Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign

Peter Gehrke iD, Philipp Klose iD, Maria Julia Pietruska iD, Günter Dhom, Octavio Weinhold, Jörg Neugebauer, Paul Weigl, Robert Sader

Dentistry Journal · 16.07.2026 · 14 (7) · 446

DOI: 10.3390/dj14070446 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentistry Journal, online erschienen am 16.07.2026. Verfasst von Peter Gehrke und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department for Oral, Cranio-Maxillofacial and Facial Plastic Surgery, Johann Wolfgang Goethe University, University Hospital Frankfurt, 60596 Frankfu…. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dentsply Sirona (Sweden).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

The aim of this study was to compare insertion torque (IT) and primary stability, assessed by resonance frequency analysis (RFA), between two implant systems with a shared macrodesign but different surface morphology and titanium grade in a standardized ex vivo porcine bone model.

Methods

Sixty implants (OmniTaper EV and Xive; n = 30 per group) were inserted into standardized porcine rib bone blocks using identical osteotomy protocols. Peak insertion torque and primary stability, assessed by resonance frequency analysis (RFA), were recorded immediately after placement. Between-group differences were evaluated using appropriate statistical methods.

Results

Bone density did not differ significantly between groups. OmniTaper EV implants demonstrated significantly higher insertion torque (28.2 ± 6.5 Ncm vs. 24.2 ± 6.2 Ncm; p = 0.020) and higher ISQ values (74.1 ± 6.0 vs. 69.8 ± 6.9; p = 0.013) than Xive implants. Mixed-effects analyses confirmed both findings after adjustment for clustering. No significant correlation was found between insertion torque and ISQ.

Conclusions

Despite sharing a common implant design concept and standardized osteotomy preparation, the investigated implant systems demonstrated significant differences in insertion torque and primary stability. The findings indicate that implant-specific characteristics, including surface morphology and titanium grade, may influence implant–bone interface mechanics during insertion. However, their individual contributions cannot be distinguished within the present study design and should therefore be interpreted as the combined influence of implant-specific characteristics.

Schlagwörter

Stichwörter der Autor:innen

ISQdental implantsex vivo studyimplant surfaceinsertion torqueprimary stabilityresonance frequency analysis

Autor:innen und Institutionen

  • Department for Oral, Cranio-Maxillofacial and Facial Plastic Surgery, Johann Wolfgang Goethe University, University Hospital Frankfurt, 60596 Frankfurt, Germany.
  • Academic Teaching and Research Institution of the Johann Wolfgang Goethe University Frankfurt am Main, Private Practice for Oral Surgery and Implant Dentistry, 67059 Ludwigshafen, Germany.
  • Private Practice, 88131 Lindau, Germany.
  • Department for Oral and Craniomaxillofacial and Plastic Surgery, University Hospital Cologne, University of Cologne Faculty of Medicine, 50931 Cologne, Germany.
  • Transfer Institute Management of Dental & Oral Medicine, Steinbeis University, 39104 Magdeburg, Germany.
  • Department of Postgraduate Education, Center for Dentistry and Oral Medicine (Carolinum), University Hospital, Goethe University Frankfurt, 60596 Frankfurt, Germany.

ORCID-Profile: Peter Gehrke, Philipp Klose, Maria Julia Pietruska

Förderung

  • Dentsply Sirona (Sweden)

Zitieren

APA

Gehrke, P., Klose, P., Pietruska, M. J., Dhom, G., Weinhold, O., Neugebauer, J., Weigl, P., & Sader, R. (2026). Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign. Dentistry Journal, 14(7), 446. https://doi.org/10.3390/dj14070446

Vancouver

Gehrke P, Klose P, Pietruska MJ, Dhom G, Weinhold O, Neugebauer J, et al. Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign. Dentistry Journal. 2026;14(7):446. doi:10.3390/dj14070446

BibTeX
@article{Gehrke2026,
  author = {Gehrke, Peter and Klose, Philipp and Pietruska, Maria Julia and Dhom, Günter and Weinhold, Octavio and Neugebauer, Jörg and Weigl, Paul and Sader, Robert},
  title = {Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign},
  journal = {Dentistry Journal},
  year = {2026},
  volume = {14},
  number = {7},
  pages = {446},
  doi = {10.3390/dj14070446}
}
Kostenlos starten →