Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign
Dentistry Journal · 16.07.2026 · 14 (7) · 446
DOI: 10.3390/dj14070446 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentistry Journal, online erschienen am 16.07.2026. Verfasst von Peter Gehrke und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department for Oral, Cranio-Maxillofacial and Facial Plastic Surgery, Johann Wolfgang Goethe University, University Hospital Frankfurt, 60596 Frankfu…. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dentsply Sirona (Sweden).
Abstract (Originalsprache)
Objectives
The aim of this study was to compare insertion torque (IT) and primary stability, assessed by resonance frequency analysis (RFA), between two implant systems with a shared macrodesign but different surface morphology and titanium grade in a standardized ex vivo porcine bone model.
Methods
Sixty implants (OmniTaper EV and Xive; n = 30 per group) were inserted into standardized porcine rib bone blocks using identical osteotomy protocols. Peak insertion torque and primary stability, assessed by resonance frequency analysis (RFA), were recorded immediately after placement. Between-group differences were evaluated using appropriate statistical methods.
Results
Bone density did not differ significantly between groups. OmniTaper EV implants demonstrated significantly higher insertion torque (28.2 ± 6.5 Ncm vs. 24.2 ± 6.2 Ncm; p = 0.020) and higher ISQ values (74.1 ± 6.0 vs. 69.8 ± 6.9; p = 0.013) than Xive implants. Mixed-effects analyses confirmed both findings after adjustment for clustering. No significant correlation was found between insertion torque and ISQ.
Conclusions
Despite sharing a common implant design concept and standardized osteotomy preparation, the investigated implant systems demonstrated significant differences in insertion torque and primary stability. The findings indicate that implant-specific characteristics, including surface morphology and titanium grade, may influence implant–bone interface mechanics during insertion. However, their individual contributions cannot be distinguished within the present study design and should therefore be interpreted as the combined influence of implant-specific characteristics.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department for Oral, Cranio-Maxillofacial and Facial Plastic Surgery, Johann Wolfgang Goethe University, University Hospital Frankfurt, 60596 Frankfurt, Germany.
- Academic Teaching and Research Institution of the Johann Wolfgang Goethe University Frankfurt am Main, Private Practice for Oral Surgery and Implant Dentistry, 67059 Ludwigshafen, Germany.
- Private Practice, 88131 Lindau, Germany.
- Department for Oral and Craniomaxillofacial and Plastic Surgery, University Hospital Cologne, University of Cologne Faculty of Medicine, 50931 Cologne, Germany.
- Transfer Institute Management of Dental & Oral Medicine, Steinbeis University, 39104 Magdeburg, Germany.
- Department of Postgraduate Education, Center for Dentistry and Oral Medicine (Carolinum), University Hospital, Goethe University Frankfurt, 60596 Frankfurt, Germany.
ORCID-Profile: Peter Gehrke, Philipp Klose, Maria Julia Pietruska
Förderung
- Dentsply Sirona (Sweden)
Zitieren
APA
Gehrke, P., Klose, P., Pietruska, M. J., Dhom, G., Weinhold, O., Neugebauer, J., Weigl, P., & Sader, R. (2026). Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign. Dentistry Journal, 14(7), 446. https://doi.org/10.3390/dj14070446
Vancouver
Gehrke P, Klose P, Pietruska MJ, Dhom G, Weinhold O, Neugebauer J, et al. Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign. Dentistry Journal. 2026;14(7):446. doi:10.3390/dj14070446
BibTeX
@article{Gehrke2026,
author = {Gehrke, Peter and Klose, Philipp and Pietruska, Maria Julia and Dhom, Günter and Weinhold, Octavio and Neugebauer, Jörg and Weigl, Paul and Sader, Robert},
title = {Influence of Implant-Specific Characteristics on Insertion Torque and Primary Stability: An Ex Vivo Comparison of Two Implant Systems with a Shared Macrodesign},
journal = {Dentistry Journal},
year = {2026},
volume = {14},
number = {7},
pages = {446},
doi = {10.3390/dj14070446}
}
EUVIMED