Temperature-Dependent Effects on Cyclic Fatigue Resistance in Three Reciprocating Endodontic Systems: An In Vitro Study
Materials · 21.02.2025 · 18 (5) · 952
DOI: 10.3390/ma18050952 ↗Kurz gefasst
Fachartikel in der Zeitschrift Materials, online erschienen am 21.02.2025. Verfasst von Marcela Salamanca Ramos und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endodontics, Rey Juan Carlos University, 28032 Madrid, Spain. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This study aimed to analyze the effect of 1% sodium hypochlorite (NaOCl) solution at different temperatures on endodontic file resistance to cyclic fatigue. A total of 90 files, Reciproc NiTi M-Wire® (REC) (n = 30), WaveOne Gold® (WOG) (n = 30), and Reciproc Blue® (RB) (n = 30) were activated under constant irrigation with 1% NaOCl at 5, 37, and 60 °C in a stainless-steel artificial canal (curvature angle = 60°; radius = 5 mm). The time to the fracture and the maximum and minimum load were recorded for each instrument, and data were subjected to statistical analysis. A higher number of cycles to fracture at 5 °C was observed between WOG and RB compared to the REC system (p < 0.05). The RB files were more resistant to cyclic fatigue fracture at 60 °C than WOG and REC (p < 0.05). No statistically significant differences were found between the mean values of the three file types used at 37 °C. The high temperature of NaOCl significantly affects the lifespan of endodontic files, making them more prone to fractures due to cyclic fatigue. The files manufactured with heat treatment showed a longer life than M-wire reciproc files.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Endodontics, Rey Juan Carlos University, 28032 Madrid, Spain.
- Unit of Endodontics and Restorative Dentistry, Department of Medical Biotechnologies, University of Siena, 53100 Siena, Italy.
- Nova Southeastern University College of Dental Medicine, Davie, FL 33314, USA.
- Surpreendente Research Group, Vila Nova de Gaia 4400-239, Portugal.
ORCID-Profile: José Aranguren, Giulia Malvicini, Carmen Bonilla, Alejandro R Perez
Zitieren
APA
Salamanca Ramos, M., Aranguren, J., Malvicini, G., De Gregorio, C., Bonilla, C., & Perez, A. R. (2025). Temperature-Dependent Effects on Cyclic Fatigue Resistance in Three Reciprocating Endodontic Systems: An In Vitro Study. Materials, 18(5), 952. https://doi.org/10.3390/ma18050952
Vancouver
Salamanca Ramos M, Aranguren J, Malvicini G, De Gregorio C, Bonilla C, Perez AR. Temperature-Dependent Effects on Cyclic Fatigue Resistance in Three Reciprocating Endodontic Systems: An In Vitro Study. Materials. 2025;18(5):952. doi:10.3390/ma18050952
BibTeX
@article{SalamancaRamos2025,
author = {Salamanca Ramos, Marcela and Aranguren, José and Malvicini, Giulia and De Gregorio, Cesar and Bonilla, Carmen and Perez, Alejandro R},
title = {Temperature-Dependent Effects on Cyclic Fatigue Resistance in Three Reciprocating Endodontic Systems: An In Vitro Study},
journal = {Materials},
year = {2025},
volume = {18},
number = {5},
pages = {952},
doi = {10.3390/ma18050952}
}