Effects of Vibrating Mesh Versus Jet Nebulization on Ventilatory Pressures During Non-Invasive Ventilation: An In Vitro Study
Life · 19.11.2026 · 16 (11) · 1957
DOI: 10.3390/life16111957 ↗Kurz gefasst
Fachartikel in der Zeitschrift Life, online erschienen am 19.11.2026. Verfasst von David Wood und 5 weiteren Autor:innen. Die Arbeit nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universidad Finis Terrae, Clínica MEDS.
Abstract (Originalsprache)
The integration of aerosol therapy with non-invasive ventilation (NIV) is a widespread clinical practice, although its effects on key ventilatory pressures remain insufficiently characterized. Vibrating mesh nebulizers (VMNs) provide superior drug delivery efficiency; however, their influence on inspiratory and expiratory positive airway pressures (IPAP and EPAP) compared to jet nebulizers (JNs) is not well defined. This study aimed to evaluate the effects of VMNs and JNs on IPAP and EPAP during NIV using a controlled in vitro model. An in vitro experimental study was performed employing a 1 L test lung coupled to a BIPAP Synchrony ventilator. Three experimental conditions were investigated: control (no nebulization), VMN (Aerogen® Solo), and JN (Hudson RCI® Micro Mist®). A total of 11 repeated measurements were obtained per condition over a 5 min period, during which proximal and distal pressures were continuously recorded. Data were analyzed via two-way repeated-measure ANOVA, with Cohen’s d employed to quantify effect sizes. Jet nebulization was associated with significant increases in EPAP at both proximal (8.12 ± 0.1 vs. 7.9 ± 0.0 cmH2O; p < 0.001) and distal measurement sites (7.7 ± 0.1 vs. 7.5 ± 0.0 cmH2O; p < 0.001), exhibiting very large effect sizes (d = 3.95–4.80). In contrast, VMN maintained EPAP at levels comparable to those of the control (d = 0.25, p = 0.96). IPAP values remained stable across all tested conditions, with no statistically significant differences observed. In this in vitro model, jet nebulizers were associated with higher expiratory airway pressures during NIV, whereas vibrating mesh nebulizers preserved pressure values comparable to those of control conditions. These findings suggest that nebulizer technology may influence pressure behavior during NIV under the experimental conditions evaluated. However, the absolute pressure differences observed were small (approximately 0.23–0.24 cmH2O) and are unlikely to be clinically relevant. Further investigation in clinical settings is warranted to determine whether these mechanical differences translate into clinically meaningful outcomes.
Förderung
- Universidad Finis Terrae
- Clínica MEDS
Zitieren
APA
Wood, D., Stöwhas, K., Aguilar, D., Aguilera, C., Jorquera, C., & Droppelmann, G. (2026). Effects of Vibrating Mesh Versus Jet Nebulization on Ventilatory Pressures During Non-Invasive Ventilation: An In Vitro Study. Life, 16(11), 1957. https://doi.org/10.3390/life16111957
Vancouver
Wood D, Stöwhas K, Aguilar D, Aguilera C, Jorquera C, Droppelmann G. Effects of Vibrating Mesh Versus Jet Nebulization on Ventilatory Pressures During Non-Invasive Ventilation: An In Vitro Study. Life. 2026;16(11):1957. doi:10.3390/life16111957
BibTeX
@article{Wood2026,
author = {Wood, David and Stöwhas, Katherine and Aguilar, Daniela and Aguilera, Camila and Jorquera, Carlos and Droppelmann, Guillermo},
title = {Effects of Vibrating Mesh Versus Jet Nebulization on Ventilatory Pressures During Non-Invasive Ventilation: An In Vitro Study},
journal = {Life},
year = {2026},
volume = {16},
number = {11},
pages = {1957},
doi = {10.3390/life16111957}
}
EUVIMED