Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison
Pharmaceutics · 06.05.2026 · 18 (5) · 575
DOI: 10.3390/pharmaceutics18050575 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pharmaceutics, online erschienen am 06.05.2026. Verfasst von Ronan MacLoughlin und 3 weiteren Autor:innen. Beteiligte Einrichtung: Emerging Technologies, Aerogen Ltd., IDA Business Park, Dangan, H91 HE94 Galway, Ireland. Die Arbeit nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The aim of this study was to assess in vitro nebulized drug delivery during invasive and non-invasive ventilation, comparing jet nebulizers (JN) and vibrating mesh nebulizers (VMN) across various pediatric ventilation models.
Methods
Drug delivery performance was compared between a continuous output JN (Aquineb) and VMN (Aerogen Solo A-VMN). The non-invasive model simulated a spontaneously breathing 9-month-old child using an anatomically correct upper airway model and breathing simulator. The invasive model used a mechanical ventilator with heated humidifier in a pediatric breathing circuit with an endotracheal tube. Nebulizers were driven with supplemental oxygen at manufacturer-recommended rates and positioned at approved locations. Absolute inhaled dose, delivery rate and residual volume were assessed using face mask, mechanical ventilation, high-flow nasal therapy and blow-by delivery methods. Dose was quantified using spectrophotometric analysis.
Results
During spontaneous breathing, A-VMN delivered almost double the dose of the evaluated JN (p < 0.001), with a significantly faster delivery rate (p < 0.001) and lower residual volume (p < 0.0001). During mechanical ventilation, A-VMN demonstrated a greater than 3-fold increase in delivered dose (p < 0.0001) and faster delivery (p < 0.0001), with reduced residual volume (p < 0.001). During high-flow nasal therapy, delivery via nasal cannula was affected by gas flow rate for both devices, with A-VMN consistently delivering greater doses. A-VMN delivered significantly greater salbutamol doses during blow-by delivery.
Conclusions
VMN demonstrated significantly superior dose delivery, faster delivery rates and reduced residual volumes compared to the evaluated JN across all tested pediatric respiratory support modalities. These in vitro findings provide important performance data for evidence-based device selection and warrant clinical investigation to determine potential therapeutic benefits in pediatric populations requiring aerosol therapy during respiratory support.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Emerging Technologies, Aerogen Ltd., IDA Business Park, Dangan, H91 HE94 Galway, Ireland.
- School of Pharmacy and Biomolecular Sciences, Royal College of Surgeons, D02 YN77 Dublin, Ireland.
- School of Pharmacy and Pharmaceutical Sciences, Trinity College, D02 PN40 Dublin, Ireland.
- Pediatric Intensive Care Unit, Assistance Publique des Hôpitaux de Paris, Hôpital Necker Enfants Malades, Université Paris Cité, 75015 Paris, France.
- Discipline of Anaesthesia, School of Medicine, Clinical Sciences Institute, University of Galway, H91 YR71 Galway, Ireland.
- Discipline of Physiology, School of Pharmacy and Medical Sciences, Human Biology Building, University of Galway, H91 W5P7 Galway, Ireland.
ORCID-Profile: Ronan MacLoughlin, Brendan D Higgins
Zitieren
APA
MacLoughlin, R., Crowe, A. M., Scully, M., & Higgins, B. D. (2026). Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison. Pharmaceutics, 18(5), 575. https://doi.org/10.3390/pharmaceutics18050575
Vancouver
MacLoughlin R, Crowe AM, Scully M, Higgins BD. Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison. Pharmaceutics. 2026;18(5):575. doi:10.3390/pharmaceutics18050575
BibTeX
@article{MacLoughlin2026,
author = {MacLoughlin, Ronan and Crowe, Ann-Marie and Scully, Michael and Higgins, Brendan D},
title = {Vibrating Mesh and Jet Nebulizer Performance in Pediatric Respiratory Support: A Multi-Modality In Vitro Comparison},
journal = {Pharmaceutics},
year = {2026},
volume = {18},
number = {5},
pages = {575},
doi = {10.3390/pharmaceutics18050575}
}
EUVIMED