EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Evaluation of the Performance of OTOPLAN-Based Cochlear Implant Electrode Array Selection: A Retrospective Study

Dayse Távora-Vieira iD, Marcus Voola, Jafri Kuthubutheen, Peter Friedland iD, Daren Gibson, Aanand Acharya

Journal of Personalized Medicine · 20.08.2023 · 13 (8) · 1276

DOI: 10.3390/jpm13081276 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Personalized Medicine, online erschienen am 20.08.2023. Verfasst von Dayse Távora-Vieira und 5 weiteren Autor:innen. Beteiligte Einrichtung: Fiona Stanley Fremantle Hospitals Group, Perth, WA 6150, Australia. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Otoplan is a surgical planning software designed to assist with cochlear implant surgery. One of its outputs is a recommendation of electrode array type based on imaging parameters. In this retrospective study, we evaluated the differences in auditory outcomes between patients who were implanted with arrays corresponding to those recommended by the Otoplan software versus those in which the array selection differed from the Otoplan recommendation. Pre-operative CT images from 114 patients were imported into the software, and array recommendations were generated. These were compared to the arrays which had actually been implanted during surgery, both in terms of array type and length. As recommended, 47% of patients received the same array, 34% received a shorter array, and 18% received a longer array. For reasons relating to structure and hearing preservation, 83% received the more flexible arrays. Those who received stiffer arrays had cochlear malformations or ossification. A negative, although non-statistically significant correlation was observed between the CNC scores at 12 months and the absolute value of the difference between recommended array and implanted array. In conclusion, clinicians may be slightly biased toward shorter electrode arrays due to their perceived greater ability to achieve full insertion. Using 3D imaging during the pre-operative planning may improve clinicians’ confidence to implant longer electrode arrays, where appropriate, to achieve optimum hearing outcomes.

Schlagwörter

Stichwörter der Autor:innen

OTOPLANanatomy-based fittingcochlear implantelectrode arrayhearing outcomespre-operative planning

Autor:innen und Institutionen

  • Fiona Stanley Fremantle Hospitals Group, Perth, WA 6150, Australia.
  • Medical School, Division of Surgery, The University of Western Australia, Perth, WA 6010, Australia.
  • Faculty of Health Sciences, School of Allied Health, Curtin University, Perth, WA 6102, Australia.
  • Sir Charles Gairdner Hospital, Perth, WA 6009, Australia.

ORCID-Profile: Dayse Távora-Vieira, Peter Friedland

Zitieren

APA

Távora-Vieira, D., Voola, M., Kuthubutheen, J., Friedland, P., Gibson, D., & Acharya, A. (2023). Evaluation of the Performance of OTOPLAN-Based Cochlear Implant Electrode Array Selection: A Retrospective Study. Journal of Personalized Medicine, 13(8), 1276. https://doi.org/10.3390/jpm13081276

Vancouver

Távora-Vieira D, Voola M, Kuthubutheen J, Friedland P, Gibson D, Acharya A. Evaluation of the Performance of OTOPLAN-Based Cochlear Implant Electrode Array Selection: A Retrospective Study. Journal of Personalized Medicine. 2023;13(8):1276. doi:10.3390/jpm13081276

BibTeX
@article{TvoraVieira2023,
  author = {Távora-Vieira, Dayse and Voola, Marcus and Kuthubutheen, Jafri and Friedland, Peter and Gibson, Daren and Acharya, Aanand},
  title = {Evaluation of the Performance of OTOPLAN-Based Cochlear Implant Electrode Array Selection: A Retrospective Study},
  journal = {Journal of Personalized Medicine},
  year = {2023},
  volume = {13},
  number = {8},
  pages = {1276},
  doi = {10.3390/jpm13081276}
}
Kostenlos starten →