Validation of Automatic Cochlear Measurements Using OTOPLAN® Software
Journal of Personalized Medicine · 08.05.2023 · 13 (5) · 805
DOI: 10.3390/jpm13050805 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Personalized Medicine, online erschienen am 08.05.2023. Verfasst von Dimitrios Paouris und 2 weiteren Autor:innen. Beteiligte Einrichtung: Clinic of Pediatric Otorhinolaryngology of the Medical Faculty, National Institute of Children's Diseases, Comenius University, 83340 Bratislava, Slo…. Die Arbeit wurde bisher 26-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Electrode length selection based on case-related cochlear parameters is becoming a standard pre-operative step for cochlear implantation. The manual measurement of the parameters is often time-consuming and may lead to inconsistencies. Our work aimed to evaluate a novel, automatic measurement method. Materials and Methods: A retrospective evaluation of pre-operative HRCT images of 109 ears (56 patients) was conducted, using a development version of the OTOPLAN® software. Inter-rater (intraclass) reliability and execution time were assessed for manual (surgeons R1 and R2) vs. automatic (AUTO) results. The analysis included A-Value (Diameter), B-Value (Width), H-Value (Height), and CDLOC-length (Cochlear Duct Length at Organ of Corti/Basilar membrane).
Results
The measurement time was reduced from approximately 7 min ± 2 (min) (manual) to 1 min (AUTO). Cochlear parameters in mm (mean ± SD) for R1, R2 and AUTO, respectively, were A-value: 9.00 ± 0.40, 8.98 ± 0.40 and 9.16 ± 0.36; B-value: 6.81 ± 0.34, 6.71 ± 0.35 and 6.70 ± 0.40; H-value: 3.98 ± 0.25, 3.85 ± 0.25 and 3.76 ± 0.22; and the mean CDLoc-length: 35.64 ± 1.70, 35.20 ± 1.71 and 35.47 ± 1.87. AUTO CDLOC measurements were not significantly different compared to R1 and R2 (H0: Rx CDLOC = AUTO CDLOC: p = 0.831, p = 0.242, respectively), and the calculated intraclass correlation coefficient (ICC) for CDLOC was 0.9 (95% CI: 0.85, 0.932) for R1 vs. AUTO; 0.90 (95% CI: 0.85, 0.932) for R2 vs. AUTO; and 0.893 (95% CI: 0.809, 0.935) for R1 vs. R2.
Conclusions
We observed excellent inter-rater reliability, a high agreement of outcomes, and reduced execution time using the AUTO method.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinic of Pediatric Otorhinolaryngology of the Medical Faculty, National Institute of Children's Diseases, Comenius University, 83340 Bratislava, Slovakia.
ORCID-Profile: Dimitrios Paouris
Zitieren
APA
Paouris, D., Kunzo, S., & Goljerová, I. (2023). Validation of Automatic Cochlear Measurements Using OTOPLAN® Software. Journal of Personalized Medicine, 13(5), 805. https://doi.org/10.3390/jpm13050805
Vancouver
Paouris D, Kunzo S, Goljerová I. Validation of Automatic Cochlear Measurements Using OTOPLAN® Software. Journal of Personalized Medicine. 2023;13(5):805. doi:10.3390/jpm13050805
BibTeX
@article{Paouris2023,
author = {Paouris, Dimitrios and Kunzo, Samuel and Goljerová, Irina},
title = {Validation of Automatic Cochlear Measurements Using OTOPLAN® Software},
journal = {Journal of Personalized Medicine},
year = {2023},
volume = {13},
number = {5},
pages = {805},
doi = {10.3390/jpm13050805}
}