The Influence of Insertion Torque on Stress Distribution in Peri-Implant Bones Around Ultra-Short Implants: An FEA Study

Mario Ceddia iD, Lorenzo Montesani, Luca Comuzzi iD, Alessandro Cipollina iD, Douglas A Deporter, Natalia Di Pietro iD, Bartolomeo Trentadue iD

Journal of Functional Biomaterials · 14.07.2025 · 16 (7) · 260

DOI: 10.3390/jfb16070260 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Functional Biomaterials, online erschienen am 14.07.2025. Verfasst von Mario Ceddia und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanics, Mathematics and Management, Polytechnic University of Bari, 70125 Bari, Italy. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 103 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Using ultra-short dental implants is a promising alternative to extensive bone grafting procedures for patients with atrophic posterior mandibles and vertical bone loss. However, the amount of insertion torque (IT) applied during implant placement significantly influences stress distribution in the peri-implant bone, which affects implant stability and long-term success. Materials and Methods: This study used finite element analysis (FEA) to examine how different insertion torques (35 N·cm and 75 N·cm) affect stress distribution in cortical and trabecular bone types D2 and D4 surrounding ultra-short implants. Von Mises equivalent stress values were compared with ultimate bone strength thresholds to evaluate the potential for microdamage during insertion.

Results

The findings demonstrate that increasing IT from 35 N·cm to 75 N·cm led to a significant increase in peri-implant bone stress. Specifically, cortical bone stress in D4 bone increased from approximately 79 MPa to 142 MPa with higher IT, exceeding physiological limits and elevating the risk of microfractures and bone necrosis. In contrast, lower IT values kept stress within safe limits, ensuring optimal primary stability without damaging the bone. These results underscore the need to strike a balance between achieving sufficient implant stability and avoiding mechanical trauma to the surrounding bone.

Conclusions

Accurate control of insertion torque during the placement of ultra-short dental implants is crucial to minimize bone damage and promote optimal osseointegration. Excessive torque, especially in low-density bone, can compromise implant success by inducing excessive stress, thereby increasing the risk of early failure.

Schlagwörter

Stichwörter der Autor:innen

atrophic jawsfinite element analysisinsertion torqueultra-short dental implants

Autor:innen und Institutionen

  • Department of Mechanics, Mathematics and Management, Polytechnic University of Bari, 70125 Bari, Italy.
  • Independent Researcher, 00131 Rome, Italy.
  • Independent Researcher, 31020 San Vendemiano, Italy.
  • Independent Researcher, 92019 Sciacca, Italy.
  • Faculty of Dentistry, University of Toronto, Toronto, ON M5G 1G6, Canada.
  • Department of Medical, Oral and Biotechnological Sciences, University of Chieti-Pescara, 66100 Chieti, Italy.

ORCID-Profile: Mario Ceddia, Luca Comuzzi, Alessandro Cipollina, Natalia Di Pietro, Bartolomeo Trentadue

Zitieren

APA

Ceddia, M., Montesani, L., Comuzzi, L., Cipollina, A., Deporter, D. A., Di Pietro, N., & Trentadue, B. (2025). The Influence of Insertion Torque on Stress Distribution in Peri-Implant Bones Around Ultra-Short Implants: An FEA Study. Journal of Functional Biomaterials, 16(7), 260. https://doi.org/10.3390/jfb16070260

Vancouver

Ceddia M, Montesani L, Comuzzi L, Cipollina A, Deporter DA, Di Pietro N, et al. The Influence of Insertion Torque on Stress Distribution in Peri-Implant Bones Around Ultra-Short Implants: An FEA Study. Journal of Functional Biomaterials. 2025;16(7):260. doi:10.3390/jfb16070260

BibTeX
@article{Ceddia2025,
  author = {Ceddia, Mario and Montesani, Lorenzo and Comuzzi, Luca and Cipollina, Alessandro and Deporter, Douglas A and Di Pietro, Natalia and Trentadue, Bartolomeo},
  title = {The Influence of Insertion Torque on Stress Distribution in Peri-Implant Bones Around Ultra-Short Implants: An FEA Study},
  journal = {Journal of Functional Biomaterials},
  year = {2025},
  volume = {16},
  number = {7},
  pages = {260},
  doi = {10.3390/jfb16070260}
}
Kostenlos starten →