Finite Element Assessment of an Implant-Integrated Transfemoral Prosthesis: Reducing Stump–Socket Stress, Pressure and Shear During Walking and Stair Negotiation
Journal of Clinical Medicine · 13.11.2026 · 15 (22) · 8882
DOI: 10.3390/jcm15228882 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 13.11.2026. Verfasst von Guerchouh Nawel und 9 weiteren Autor:innen. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Recep Tayyip Erdoğan University Development Foundation.
Abstract (Originalsprache)
Background/Objectives: Lower-limb amputation presents biomechanical challenges at the residual limb–prosthetic socket interface, where uneven load transfer may generate localized stress concentrations. This study investigates the biomechanical behavior of a patient-specific transfemoral residual limb–prosthesis system with and without a structural implant positioned at the distal femur.
Methods
Finite element simulations were performed for three activities: normal walking, stair descent, and stair ascent. Von Mises stress, contact pressure (CPRESS), and longitudinal shear stress (CSHEAR1) were evaluated in the residual limb and prosthetic components.
Results
During normal walking, stair descent, and stair ascent, the implant reduced the maximum von Mises stress in the residual limb by 62.9%, 55.1%, and 80.2%, respectively. The maximum longitudinal shear stress in the residual limb decreased by 28.6%, 28.0%, and 41.5%, respectively. During stair descent, the maximum contact pressure at the residual limb interface decreased by 79.4%, from 173 to 35.6 kPa, while during stair ascent it decreased by 38.6%, from 95.5 to 58.6 kPa. The maximum contact pressure on the prosthetic liner decreased by 43.0% during stair descent and 33.7% during stair ascent. Similarly, the maximum longitudinal shear stress on the socket decreased by 43.3% and 33.3%, respectively. However, slight increases in some peak contact and shear stress values were observed during normal walking. These findings indicate that the biomechanical effects of the implant depend on the activity and the parameter evaluated.
Conclusions
Although the numerical results demonstrate reductions in several peak stress and pressure values, clinical benefits cannot be established from the simulations alone.
Förderung
- Recep Tayyip Erdoğan University Development Foundation
Zitieren
APA
Nawel, G., Abdelmadjid, M., Sallah, Z. M. E., Abderahmane, S., Ismail, B., Yaylacı, M., Öztürk, Ş., Yaylacı, E. U., Akdoğan, A., & Güvercin, Y. (2026). Finite Element Assessment of an Implant-Integrated Transfemoral Prosthesis: Reducing Stump–Socket Stress, Pressure and Shear During Walking and Stair Negotiation. Journal of Clinical Medicine, 15(22), 8882. https://doi.org/10.3390/jcm15228882
Vancouver
Nawel G, Abdelmadjid M, Sallah ZME, Abderahmane S, Ismail B, Yaylacı M, et al. Finite Element Assessment of an Implant-Integrated Transfemoral Prosthesis: Reducing Stump–Socket Stress, Pressure and Shear During Walking and Stair Negotiation. Journal of Clinical Medicine. 2026;15(22):8882. doi:10.3390/jcm15228882
BibTeX
@article{Nawel2026,
author = {Nawel, Guerchouh and Abdelmadjid, Moulgada and Sallah, Zagane Mohammed El and Abderahmane, Sahli and Ismail, Boudjemaa and Yaylacı, Murat and Öztürk, Şevval and Yaylacı, Ecren Uzun and Akdoğan, Ali and Güvercin, Yılmaz},
title = {Finite Element Assessment of an Implant-Integrated Transfemoral Prosthesis: Reducing Stump–Socket Stress, Pressure and Shear During Walking and Stair Negotiation},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {22},
pages = {8882},
doi = {10.3390/jcm15228882}
}
EUVIMED