Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study
Journal of Clinical Medicine · 12.11.2026 · 15 (22) · 8846
DOI: 10.3390/jcm15228846 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Journal of Clinical Medicine, online erschienen am 12.11.2026. Verfasst von Mariusz Stanisław Sowa und 2 weiteren Autor:innen. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Minister of Science under the ‘Regional Initiative of Excellence Program’.
Abstract (Originalsprache)
Background
This study aims to establish local diagnostic reference levels (DRLs) for mechanical thrombectomy performed using a low-dose imaging protocol and to assess objective image quality parameters under routine clinical conditions.
Methods
This retrospective, single-center observational study included 59 consecutive patients who underwent mechanical thrombectomy for acute ischemic stroke between January 2023 and March 2026, with no exclusions. Radiation metrics, including dose-area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and number of images, were extracted from system-generated dose reports. Local DRLs were defined as the 75th percentile (P75) of the dose distribution. Objective image quality was assessed using quantitative parameters such as background noise, signal-to-noise ratio (SNR), absolute contrast-to-noise ratio (CNR), CNR/DAP, CNR/Ka,r, and figure-of-merit (FOM = CNR2/DAP).
Results
In the overall cohort, the median and P75 values were 16.6/26.2 Gy·cm2 for DAP, 206/306 mGy for Ka,r, and 1217/1622 s for FT. The median number of images per procedure was 210, with a P75 of 285. Procedures utilizing combined biplane imaging demonstrated higher DAP and Ka,r values compared to those using single-plane imaging, while FT did not differ significantly. Objective image-quality analysis was feasible in all 59 procedures. The median background noise, SNR, and absolute CNR were 44.24, 34.47, and 26.99, respectively. Median dose-normalized values were 1.32 for CNR/DAP, 0.12 for CNR/Ka,r, and 33.21 for FOM. No procedure required escalation of the imaging protocol due to insufficient angiographic visualization.
Conclusions
Mechanical thrombectomy performed under a predefined low-dose protocol yielded low absolute local dose values in this single-center cohort. ROI-based image-quality parameters remained measurable. Because no standard-dose comparator was imaged, the magnitude of dose reduction and the preservation of image quality relative to conventional settings could not be quantified.
Förderung
- Minister of Science under the ‘Regional Initiative of Excellence Program’
Zitieren
APA
Sowa, M. S., Sowa, J., & Budzanowski, M. (2026). Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study. Journal of Clinical Medicine, 15(22), 8846. https://doi.org/10.3390/jcm15228846
Vancouver
Sowa MS, Sowa J, Budzanowski M. Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study. Journal of Clinical Medicine. 2026;15(22):8846. doi:10.3390/jcm15228846
BibTeX
@article{Sowa2026,
author = {Sowa, Mariusz Stanisław and Sowa, Joanna and Budzanowski, Maciej},
title = {Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {22},
pages = {8846},
doi = {10.3390/jcm15228846}
}
EUVIMED