Stress Distribution in Interlocking and Blocking Screw Fixation for Distal Tibial Intramedullary Nailing: A Finite Element Analysis
Journal of Clinical Medicine · 05.07.2025 · 14 (13) · 4769
DOI: 10.3390/jcm14134769 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 05.07.2025. Verfasst von Gu-Hee Jung und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Gyeongsang National University, College of Medicine, Gyeongsang National University Changwon Hospital, Samjeongja-…. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug Safety, Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Mi….
Abstract (Originalsprache)
Background/Objectives: This study evaluates the structural strength of various fixation models for distal tibial fractures using finite element analysis, comparing the biomechanical performance of different interlocking and blocking screw configurations in intramedullary nail fixation.
Methods
Finite element models were developed for three different screw configurations: two interlocking screws (T2S), three interlocking screws (T3S), and two interlocking screws combined with two blocking screws (T2SBS2). Material properties were assigned using SAWBONES® models, reflecting established biomechanical characteristics. The models were subjected to static axial loading (800 Nm), internal rotation (3500 Nm), and external rotation stresses. The stress distribution and structural integrity of each fixation model were analyzed.
Results
The T3S model demonstrated the lowest maximum von Mises stress values across all loading conditions. Under axial loading, the maximum VMS at the medial first screw contact was the lowest in the T3S model (65.96 MPa) compared to T2S (135.68 MPa) and T2SBS2 (150.30 MPa). Similarly, internal rotation loading resulted in the lowest stress at the medial first screw site in T3S (64.86 MPa), compared to significantly higher stresses in T2S (136.29 MPa) and T2SBS2 (158.11 MPa). The T2SBS2 configuration showed effective stress reduction at the distal interlocking screws, highlighting its potential for improved load sharing.
Conclusions
The findings underscore the importance of the secure fixation of the initial interlocking screw in distal tibial fractures. The fixation model utilizing three interlocking screws provided the most favorable overall stress distribution, whereas the inclusion of blocking screws effectively reduced stress concentrations at distal screw sites.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedic Surgery, Gyeongsang National University, College of Medicine, Gyeongsang National University Changwon Hospital, Samjeongja-ro 1, Seongsan-gu, Changwon-si 51472, Republic of Korea.
- Institute of Health Sciences (Medical ICT Convergence Research Center), College of Medicine, Gyeongsang National University, Jinju 52727, Republic of Korea.
ORCID-Profile: Gu-Hee Jung, Se-Lin Jeong, Jungtae Ahn
Förderung
- the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug Safety
- Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Mi…
Zitieren
APA
Jung, G. H., Jeong, S. L., & Ahn, J. (2025). Stress Distribution in Interlocking and Blocking Screw Fixation for Distal Tibial Intramedullary Nailing: A Finite Element Analysis. Journal of Clinical Medicine, 14(13), 4769. https://doi.org/10.3390/jcm14134769
Vancouver
Jung GH, Jeong SL, Ahn J. Stress Distribution in Interlocking and Blocking Screw Fixation for Distal Tibial Intramedullary Nailing: A Finite Element Analysis. Journal of Clinical Medicine. 2025;14(13):4769. doi:10.3390/jcm14134769
BibTeX
@article{Jung2025,
author = {Jung, Gu-Hee and Jeong, Se-Lin and Ahn, Jungtae},
title = {Stress Distribution in Interlocking and Blocking Screw Fixation for Distal Tibial Intramedullary Nailing: A Finite Element Analysis},
journal = {Journal of Clinical Medicine},
year = {2025},
volume = {14},
number = {13},
pages = {4769},
doi = {10.3390/jcm14134769}
}
EUVIMED