Accuracy of a Dual-Sensor Heat-Flux (DHF) Non-Invasive Core Temperature Sensor in Pediatric Patients Undergoing Surgery
Journal of Clinical Medicine · 09.11.2023 · 12 (22) · 7018
DOI: 10.3390/jcm12227018 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 09.11.2023. Verfasst von Sebastian Zeiner und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesia, Intensive Care Medicine and Pain Medicine, Medical University of Vienna, 1090 Vienna, Austria. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Dräger (Germany), Dräger Austria GmbH, 1230 Vienna, Austria.
Abstract (Originalsprache)
Accurate temperature measurement is crucial for the perioperative management of pediatric patients, and non-invasive thermometry is necessary when invasive methods are infeasible. A prospective observational study was conducted on 57 patients undergoing elective surgery. Temperatures were measured using a dual-sensor heat-flux (DHF) thermometer (Tcore™) and a rectal temperature probe (TRec), and the agreement between the two measurements was assessed. The DHF measurements showed a bias of +0.413 °C compared with those of the TRec. The limits of agreement were broader than the pre-defined ±0.5 °C range (−0.741 °C and +1.567 °C). Although the DHF sensors tended to overestimate the core temperature compared to the rectal measurements, an error grid analysis demonstrated that 95.81% of the DHF measurements would not have led to a wrong clinical decision, e.g., warming or cooling when not necessary. In conclusion, the low number of measurements that would have led to incorrect decisions suggests that the DHF sensor can be considered an option for continuous temperature measurement when more invasive methods are infeasible.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anaesthesia, Intensive Care Medicine and Pain Medicine, Medical University of Vienna, 1090 Vienna, Austria.
- Ludwig Boltzmann Institute Digital Health and Patient Safety (LBI DHPS), 1090 Vienna, Austria.
- Institute of Clinical Biometrics, Center for Medical Data Science, Medical University of Vienna, 1090 Vienna, Austria.
ORCID-Profile: Sebastian Zeiner, Markus Zadrazil, Marion Wiegele, Peter Marhofer, Fabian Peter Hammerle, Daniel Laxar, Andreas Gleiss, Oliver Kimberger
Förderung
- Dräger (Germany)
- Dräger Austria GmbH, 1230 Vienna, Austria
Zitieren
APA
Zeiner, S., Zadrazil, M., Willschke, H., Wiegele, M., Marhofer, P., Hammerle, F. P., Laxar, D., Gleiss, A., & Kimberger, O. (2023). Accuracy of a Dual-Sensor Heat-Flux (DHF) Non-Invasive Core Temperature Sensor in Pediatric Patients Undergoing Surgery. Journal of Clinical Medicine, 12(22), 7018. https://doi.org/10.3390/jcm12227018
Vancouver
Zeiner S, Zadrazil M, Willschke H, Wiegele M, Marhofer P, Hammerle FP, et al. Accuracy of a Dual-Sensor Heat-Flux (DHF) Non-Invasive Core Temperature Sensor in Pediatric Patients Undergoing Surgery. Journal of Clinical Medicine. 2023;12(22):7018. doi:10.3390/jcm12227018
BibTeX
@article{Zeiner2023,
author = {Zeiner, Sebastian and Zadrazil, Markus and Willschke, Harald and Wiegele, Marion and Marhofer, Peter and Hammerle, Fabian Peter and Laxar, Daniel and Gleiss, Andreas and Kimberger, Oliver},
title = {Accuracy of a Dual-Sensor Heat-Flux (DHF) Non-Invasive Core Temperature Sensor in Pediatric Patients Undergoing Surgery},
journal = {Journal of Clinical Medicine},
year = {2023},
volume = {12},
number = {22},
pages = {7018},
doi = {10.3390/jcm12227018}
}
EUVIMED