Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices
Journal of Cardiovascular Development and Disease · 08.10.2024 · 11 (10) · 313
DOI: 10.3390/jcdd11100313 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Development and Disease, online erschienen am 08.10.2024. Verfasst von Hugo Lanz und 16 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine I, LMU University Hospital, 81377 Munich, Germany. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
MRI (magnetic resonance imaging) represents the diagnostic image modality of choice in several conditions. With an increasing number of patients requiring MRI for diagnostic purposes, the issue of safety in patients with cardiac implantable electronic devices (CIED) undergoing this imaging modality will play an ever more important role. The purpose of this study was to assess the safety and device function following MRI in an unrestricted real-world cohort of patients with a wide array of cardiac devices.
Methods
We conducted a retrospective single-center study including 1010 MRI studies conducted in adult patients (≥18 years) with an implanted CIED treated in the University Hospital of Munich (LMU) between July 2012 and March 2024. Patients with non-MR conditionally labeled leads, abandoned or epicardial leads, as well as lead fragments, were included for analysis.
Results
Across a total of 1010 MRIs (920 total MR-conditional device generators) performed in patients with an implanted CIED, there were no deaths, reports of discomfort, palpitations, heating, or ventricular arrythmias in the 24 h following MRI. Only 2/1010 MRIs were followed by a reported atrial arrhythmia within 24 h, both in patients with an MR-conditional pacemaker (PM) device without an abandoned lead. No significant changes in device function following MRI from baseline were observed across all included CIEDs. Lastly, no instances of severe malfunction, such as generator failure, loss of capture, electrical reset, or inappropriate inhibition of pacing, were found in post-MRI interrogation reports across all MRI studies.
Conclusions
Based on the analysis of 1010 MRIs undergone by patients with CIEDs, following standardized device interrogation, manufacturer-advised device programming, monitoring of vital function, and manufacturer-advised reprogramming, MRI can be performed safely and without adverse events or changes in device function.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medicine I, LMU University Hospital, 81377 Munich, Germany.
- DZHK (German Center for Cardiovascular Research), Partner Site Munich Heart Alliance, 80636 Munich, Germany.
- Institute for Medical Information Processing, Biometry, and Epidemiology, Ludwig-Maximilians-Universität München, 81377 Munich, Germany.
- Departments of Cardiac Surgery, LMU University Hospital, 81377 Munich, Germany.
ORCID-Profile: Julia Höpler, Marie Kraft, Leonhard Binzenhöfer, Nils Gade, Sebastian Sadoni
Zitieren
APA
Lanz, H., Strauß, K., Höpler, J., Kraft, M., Hoffmann, S., Binzenhöfer, L., Gade, N., Roden, D., Saleh, I., Kääb, S., Lackermair, K., Sadoni, S., Hagl, C., Massberg, S., Estner, H., Fichtner, S., & Lüsebrink, E. (2024). Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices. Journal of Cardiovascular Development and Disease, 11(10), 313. https://doi.org/10.3390/jcdd11100313
Vancouver
Lanz H, Strauß K, Höpler J, Kraft M, Hoffmann S, Binzenhöfer L, et al. Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices. Journal of Cardiovascular Development and Disease. 2024;11(10):313. doi:10.3390/jcdd11100313
BibTeX
@article{Lanz2024,
author = {Lanz, Hugo and Strauß, Katharina and Höpler, Julia and Kraft, Marie and Hoffmann, Sabine and Binzenhöfer, Leonhard and Gade, Nils and Roden, Daniel and Saleh, Inas and Kääb, Stefan and Lackermair, Korbinian and Sadoni, Sebastian and Hagl, Christian and Massberg, Steffen and Estner, Heidi and Fichtner, Stephanie and Lüsebrink, Enzo},
title = {Safety of Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices},
journal = {Journal of Cardiovascular Development and Disease},
year = {2024},
volume = {11},
number = {10},
pages = {313},
doi = {10.3390/jcdd11100313}
}
EUVIMED