Metrological-Characteristics-Based Calibration of Optical Areal Surface Measuring Instruments and Evaluation of Measurement Uncertainty for Surface Texture Measurements
Instruments · 08.05.2025 · 9 (2) · 11
DOI: 10.3390/instruments9020011 ↗Kurz gefasst
Fachartikel in der Zeitschrift Instruments, online erschienen am 08.05.2025. Verfasst von Sai Gao und 2 weiteren Autor:innen. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union.
Abstract (Originalsprache)
ISO 25178 part 600:2019 and part 700:2022 introduce a calibration framework based on seven metrological characteristics (MCs) for calibrating optical areal surface measuring instruments. Among these, topography fidelity is a newly defined metrological characteristic that remains a critical yet unresolved challenge in instrument calibration. This paper proposes strategies to address topography fidelity, including a key criterion for selecting suitable instrument setups by comparing slope measurement capability with local surface slopes, as well as methods for investigating the field-of-view homogeneity and directional performance difference along the x- and y-axes. Furthermore, the uncertainty contribution of topography fidelity in surface topography measurements is analysed. The paper also determines the uncertainty associated with the remaining six MCs. Based on the proposed MC-based calibration approach and the corresponding uncertainty contributions, an overall measurement uncertainty model for Sa and Sq parameters is presented. Finally, uncertainty evaluations for Sa and Sq are demonstrated on a challenging surface, where topography fidelity plays a significant role in the measurement uncertainty evaluation.
Förderung
- European Union
Zitieren
APA
Gao, S., Felgner, A., & Brand, U. (2025). Metrological-Characteristics-Based Calibration of Optical Areal Surface Measuring Instruments and Evaluation of Measurement Uncertainty for Surface Texture Measurements. Instruments, 9(2), 11. https://doi.org/10.3390/instruments9020011
Vancouver
Gao S, Felgner A, Brand U. Metrological-Characteristics-Based Calibration of Optical Areal Surface Measuring Instruments and Evaluation of Measurement Uncertainty for Surface Texture Measurements. Instruments. 2025;9(2):11. doi:10.3390/instruments9020011
BibTeX
@article{Gao2025,
author = {Gao, Sai and Felgner, André and Brand, Uwe},
title = {Metrological-Characteristics-Based Calibration of Optical Areal Surface Measuring Instruments and Evaluation of Measurement Uncertainty for Surface Texture Measurements},
journal = {Instruments},
year = {2025},
volume = {9},
number = {2},
pages = {11},
doi = {10.3390/instruments9020011}
}