Validation of a Novel Medical Device Sterilization Modality Using Nitric Oxide (NO): Sterility Efficacy and Initial Material Compatibility

Weilue He, Ethan Sommer, Megan C. Frost

International Journal of Medical Devices · 12.08.2026 · 1 (1) · 5

DOI: 10.3390/ijmd1010005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Medical Devices, online erschienen am 12.08.2026. Verfasst von Weilue He und 2 weiteren Autor:innen. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sterile State Inc..

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Terminal sterilization is a regulatory requirement for most medical devices. However, conventional sterilization methods can compromise temperature-sensitive and biologically reactive components. This study evaluates the sterilization efficacy and preliminary material compatibility of a novel room-temperature sterilization modality using gaseous nitric oxide (NO).

Methods

NO was delivered using standardized Sterile Solution® (Sterile State Inc., Grand Rapids, MI, USA). To characterize microbial inactivation kinetics, the decimal reduction time (D-value) of commercial biological indicators (BIs), Bacillus atrophaeus Apex® Discs, was determined in accordance with ISO 11138. BIs were exposed to NO at ambient temperature (18.9–22.0 °C). D-values were calculated using both the survivor curve method (Holcomb–Spearman–Karber) and the fraction-negative method (Stumbo–Murphy–Cochran). Material compatibility was evaluated across four primary medical device and packaging plastics, including polyethylene (PE), poly(ethylene terephthalate) (PET), polypropylene (PP), and polyvinyl chloride (PVC)), representing 70% of the industry market share. After exposure to NO sterilization cycles, mechanical and chemical properties were characterized using tensile testing (ASTM D882-18) and Fourier-transform infrared (FTIR) spectroscopy relative to unexposed controls.

Results

The calculated D-values ranged from 1.31 (78.7 min) to 1.34 h (80.4 min) across both analytical methods, corresponding to an estimated 6-Log reduction in Bacillus atrophaeus BIs in approximately 8 h. Meanwhile, plastics exposed to the NO sterilization cycle showed no significant changes in mechanical or chemical properties.

Conclusions

These findings demonstrate that NO sterilization achieves semi-Log-linear inactivation of a widely accepted BI and represents a promising alternative to conventional gaseous sterilization methods such as ethylene oxide (EtO). By achieving a reliable sterility assurance level under mild ambient conditions without adversely affecting the physicochemical properties of commonly used plastic materials, NO sterilization may offer a practical solution for thermosensitive and delicate medical devices.

Förderung

  • Sterile State Inc.

Zitieren

APA

He, W., Sommer, E., & Frost, M. C. (2026). Validation of a Novel Medical Device Sterilization Modality Using Nitric Oxide (NO): Sterility Efficacy and Initial Material Compatibility. International Journal of Medical Devices, 1(1), 5. https://doi.org/10.3390/ijmd1010005

Vancouver

He W, Sommer E, Frost MC. Validation of a Novel Medical Device Sterilization Modality Using Nitric Oxide (NO): Sterility Efficacy and Initial Material Compatibility. International Journal of Medical Devices. 2026;1(1):5. doi:10.3390/ijmd1010005

BibTeX
@article{He2026,
  author = {He, Weilue and Sommer, Ethan and Frost, Megan C.},
  title = {Validation of a Novel Medical Device Sterilization Modality Using Nitric Oxide (NO): Sterility Efficacy and Initial Material Compatibility},
  journal = {International Journal of Medical Devices},
  year = {2026},
  volume = {1},
  number = {1},
  pages = {5},
  doi = {10.3390/ijmd1010005}
}
Kostenlos starten →