EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Design and Implementation of a High-Resolution Real-Time Ultrasonic Endoscopy Imaging System Based on FPGA and Coded Excitation

Haihang Gu, Fujia Sun, Shuhao Hou, Shuangyuan Wang

Electronics · 06.04.2026 · 15 (7) · 1526

DOI: 10.3390/electronics15071526 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Electronics, online erschienen am 06.04.2026. Verfasst von Haihang Gu und 3 weiteren Autor:innen. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

High-frequency endoscopic ultrasound is crucial for the early diagnosis of gastrointestinal tumors. However, achieving high axial resolution, deep tissue signal-to-noise ratio, and real-time data processing simultaneously remains a significant challenge in hardware implementation. This paper proposes a miniaturized real-time high-frequency imaging system based on the Xilinx Artix-7 FPGA. To overcome attenuation limitations of high-frequency signals, we employ a 4-bit Barker code-encoded excitation scheme coupled with a programmable ±100 V high-voltage transmission circuit. This effectively enhances echo energy without exceeding peak voltage safety thresholds. At the receiver end, the system utilizes a multi-channel analog front end integrated with mixed-signal time-gain compensation technology. Furthermore, to address transmission bottlenecks for massive echo data, we designed a Low-Voltage Differential Signaling (LVDS) interface logic based on dynamic phase calibration, ensuring stable, high-speed data transfer to the host computer via USB 3.0. Experimental results with a 20 MHz transducer demonstrate that the system achieves real-time B-mode imaging at 30 frames per second. Phantom testing revealed an axial resolution of 0.13 mm, enabling clear differentiation of 0.1 mm microstructures. Compared to conventional single-pulse excitation, coded excitation technology improved signal-to-noise ratio (SNR) by approximately 4.5 dB at a depth of 40 mm. These results validate the system’s capability for high-precision deep imaging suitable for clinical endoscopy applications, delivered in a compact, low-power form factor.

Zitieren

APA

Gu, H., Sun, F., Hou, S., & Wang, S. (2026). Design and Implementation of a High-Resolution Real-Time Ultrasonic Endoscopy Imaging System Based on FPGA and Coded Excitation. Electronics, 15(7), 1526. https://doi.org/10.3390/electronics15071526

Vancouver

Gu H, Sun F, Hou S, Wang S. Design and Implementation of a High-Resolution Real-Time Ultrasonic Endoscopy Imaging System Based on FPGA and Coded Excitation. Electronics. 2026;15(7):1526. doi:10.3390/electronics15071526

BibTeX
@article{Gu2026,
  author = {Gu, Haihang and Sun, Fujia and Hou, Shuhao and Wang, Shuangyuan},
  title = {Design and Implementation of a High-Resolution Real-Time Ultrasonic Endoscopy Imaging System Based on FPGA and Coded Excitation},
  journal = {Electronics},
  year = {2026},
  volume = {15},
  number = {7},
  pages = {1526},
  doi = {10.3390/electronics15071526}
}
Kostenlos starten →