Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro Study
Dentistry Journal · 07.04.2026 · 14 (4) · 214
DOI: 10.3390/dj14040214 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentistry Journal, online erschienen am 07.04.2026. Verfasst von André Luís Onodera und 4 weiteren Autor:innen. Beteiligte Einrichtung: Implant Dentistry, São Leopoldo Mandic School of Dentistry, Campinas 13045-755, SP, Brazil. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: It remains unclear which drilling strategy is most effective for maximizing mechanical stability in low-density bone and whether high insertion torque is determinative. The aim of this in vitro study was to compare the primary stability of implants placed using different drilling protocols—conventional (CV), undersized (US), and osseodensification (OD). Three osseodensification systems—Versah burs (V), Bone Reamer Drills (WF), and Master Conical Densifiers (DSP)—were also compared.
Methods
A set of 11 blocks was used for the drilling protocol comparison (CV, US, OD) and a separate set of 11 blocks was used for the osseodensification system comparison (V, WF, DSP). External-hexagon implants were inserted epicrestally. Insertion torque was measured using a torque meter, and implant stability quotients (ISQs) were assessed through resonance frequency analysis.
Results
ISQ for OD was significantly higher than that for CV but statistically similar to that for US, whereas insertion torque for OD was significantly higher than that for both US and CV. A weak correlation was found between variables for CV and US, and a moderate one was observed for OD. Both WF and DSP showed significantly higher ISQ values than V. Insertion torque for DSP was significantly higher than that for both WF and V. A moderate correlation was found between variables for DSP and V, and a weak one for WF.
Conclusions
In this invitro study, the OD protocol performed better than CV in terms of ISQ and better than both CV and US in terms of insertion torque. WF and DSP outperformed V in ISQ, whereas DSP yielded the highest insertion torque. Weak-to-moderate correlations between variables in both analyses indicated that higher insertion torque did not necessarily translate into greater stability.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Implant Dentistry, São Leopoldo Mandic School of Dentistry, Campinas 13045-755, SP, Brazil.
- Implant Dentistry, School of Dentistry, University of Taubaté, Taubaté 12020-340, SP, Brazil.
ORCID-Profile: André Luís Onodera, Alexandre Wanderley Alécio, Gustavo Batista Grolli Klein, Rogério de Lima Romeiro
Zitieren
APA
Onodera, A. L., Alécio, A. W., Klein, G. B. G., Cortelli, S., & Romeiro, R. L. (2026). Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro Study. Dentistry Journal, 14(4), 214. https://doi.org/10.3390/dj14040214
Vancouver
Onodera AL, Alécio AW, Klein GBG, Cortelli S, Romeiro RL. Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro Study. Dentistry Journal. 2026;14(4):214. doi:10.3390/dj14040214
BibTeX
@article{Onodera2026,
author = {Onodera, André Luís and Alécio, Alexandre Wanderley and Klein, Gustavo Batista Grolli and Cortelli, Sheila and Romeiro, Rogério de Lima},
title = {Primary Stability of Implants Inserted Using Different Osseodensification Systems in Low-Density Bone: An In Vitro Study},
journal = {Dentistry Journal},
year = {2026},
volume = {14},
number = {4},
pages = {214},
doi = {10.3390/dj14040214}
}
EUVIMED