Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison
Diagnostics · 12.11.2026 · 16 (22) · 3745
DOI: 10.3390/diagnostics16223745 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 12.11.2026. Verfasst von Rodrigo Torres-Castro und 7 weiteren Autor:innen. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Maximal inspiratory (MIP) and expiratory (MEP) pressures are widely used indices of respiratory muscle strength, but clinical manometers are costly. We evaluated the bench agreement of two units of a low-cost digital manometer with a clinical respiratory pressure comparator.
Methods
Two BioPimo 101® units and a Vitalograph Pneumotrac 6800® were connected simultaneously to a mechanical insufflation–exsufflation device coupled to a test lung. Pressures of 5–70 cmH2O were applied in 5-cmH2O steps under two compliance conditions, in triplicate (56 conditions; 336 paired differences analyzed as repeated measurements). Agreement was assessed with Bland–Altman analysis, condition-clustered bootstrap confidence intervals, regression-based limits of agreement and ICC (2,1).
Results
For MIP, mean differences were −0.03 to 0.21 cmH2O (limits of agreement −0.42 to 0.72 cmH2O). For MEP, both units showed higher readings than the comparator (mean differences 0.52–0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44–2.64 cmH2O. ICCs exceeded 0.999.
Conclusions
Both units agreed closely with the comparator under bench conditions, particularly for MIP, whereas MEP showed positive, pressure-dependent differences. Without a traceable pressure standard and with only two units tested, these findings do not establish absolute accuracy or clinical interchangeability.
Zitieren
APA
Torres-Castro, R., Otto-Yáñez, M., Saldaña-Villegas, C., Vasconcello-Castillo, L., Binetti, G., Puppo, H., Barros-Poblete, M., & Vera-Uribe, R. (2026). Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison. Diagnostics, 16(22), 3745. https://doi.org/10.3390/diagnostics16223745
Vancouver
Torres-Castro R, Otto-Yáñez M, Saldaña-Villegas C, Vasconcello-Castillo L, Binetti G, Puppo H, et al. Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison. Diagnostics. 2026;16(22):3745. doi:10.3390/diagnostics16223745
BibTeX
@article{TorresCastro2026,
author = {Torres-Castro, Rodrigo and Otto-Yáñez, Matías and Saldaña-Villegas, Claudia and Vasconcello-Castillo, Luis and Binetti, Giovanna and Puppo, Homero and Barros-Poblete, Marisol and Vera-Uribe, Roberto},
title = {Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {22},
pages = {3745},
doi = {10.3390/diagnostics16223745}
}
EUVIMED