Validation of a Low‐Cost Manometer to Assess of Tongue, Lip, Cheek, and Respiratory Strength: A Laboratory‐Based Study

James A Curtis iD, Crystal Diaz, Theresa Lee, Anaïs Rameau iD

The Laryngoscope · 23.08.2024

DOI: 10.1002/lary.31727 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The Laryngoscope, online erschienen am 23.08.2024. Verfasst von James A Curtis und 3 weiteren Autor:innen. Beteiligte Einrichtung: Aerodigestive Innovations Research lab (AIR), Department of Otolaryngology-Head & Neck Surgery, Weill Cornell Medical College of Cornell University, …. Hauptthemen laut MEDLINE: Cheek, Tongue, Manometry, Lip. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: NIA NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective(s) The objective of this study was to characterize the level of agreement between three manometers: (1) Iowa Oral Performance Instrument (IOPI)—the reference standard for tongue, lip, and cheek strength assessments; (2) MicroRPM Respiratory Pressure Meter (MicroRPM)—the reference standard for respiratory strength assessments; and (3) Digital Pressure Manometer (DPM)—an alternative, low‐cost pressure testing manometer.

Methods

Manual pressures were simultaneously applied to the IOPI and DPM, and to the MicroRPM and DPM, within a controlled laboratory setting. Agreement in pressure readings were analyzed using descriptive statistics, Lin's concordance correlation, and Bland–Altman Plots. Agreement was interpreted as “poor” if ρ c < 0.90, “moderate” if ρ c = 0.90 – < 0.95, “substantial” if ρ c = 0.95 – < 0.99, and “excellent” if ρ c ≥ 0.99.

Results

Differences in pressure readings between the DPM and clinical reference standards were consistently present yet highly predictable. There was a median absolute difference of 2.0–3.9 kPa between the IOPI and DPM, and 4.5–9.8 cm H 2 O between the MicroRPM and DPM. Lin's concordance revealed “substantial” agreement between the IOPI and DPM ( ρ c = 0.98) and the MicroRPM and DPM ( ρ c = 0.99).

Conclusion

The DPM revealed higher pressure readings when compared to the IOPI and MicroRPM. However, differences in pressure readings were relatively small, highly predictable, and yielded substantial overall agreement. These findings suggest the DPM may be a valid, lower‐cost alternative for objective assessments of tongue, lip, cheek, and respiratory muscle strength. Future research should expand on the present findings in clinical patient populations. Level of Evidence NA Laryngoscope , 2024

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CheekTongueManometryLipHumansMuscle StrengthMaleReproducibility of ResultsAdultFemalePressure

Stichwörter der Autor:innen

DPMIOPIMicroRPMcoughswallowingvalidity

Autor:innen und Institutionen

  • Aerodigestive Innovations Research lab (AIR), Department of Otolaryngology-Head & Neck Surgery, Weill Cornell Medical College of Cornell University, New York, New York, U.S.A.
  • Communication Sciences and Disorders Program, Department of Biobehavioral Sciences, Teachers College of Columbia University, New York, New York, U.S.A.
  • Laryngology Innovation Lab, Department of Otolaryngology-Head & Neck Surgery, Weill Cornell Medical College of Cornell University, New York, New York, U.S.A.

ORCID-Profile: James A Curtis, Anaïs Rameau

Förderung

  • NIA NIH HHS

Zitieren

APA

Curtis, J. A., Diaz, C., Lee, T., & Rameau, A. (2024). Validation of a Low‐Cost Manometer to Assess of Tongue, Lip, Cheek, and Respiratory Strength: A Laboratory‐Based Study. The Laryngoscope. https://doi.org/10.1002/lary.31727

Vancouver

Curtis JA, Diaz C, Lee T, Rameau A. Validation of a Low‐Cost Manometer to Assess of Tongue, Lip, Cheek, and Respiratory Strength: A Laboratory‐Based Study. The Laryngoscope. 2024. doi:10.1002/lary.31727

BibTeX
@article{Curtis2024,
  author = {Curtis, James A and Diaz, Crystal and Lee, Theresa and Rameau, Anaïs},
  title = {Validation of a Low‐Cost Manometer to Assess of Tongue, Lip, Cheek, and Respiratory Strength: A Laboratory‐Based Study},
  journal = {The Laryngoscope},
  year = {2024},
  doi = {10.1002/lary.31727}
}
Kostenlos starten →