A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy
Diagnostics · 05.11.2026 · 16 (21) · 3664
DOI: 10.3390/diagnostics16213664 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 05.11.2026. Verfasst von Markus Stana und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Paracelsus Medical University, 5020 Salzburg, Austria. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Accurate characterization of treatment geometry is essential for intraoperative electron radiotherapy (IOeRT) treatment planning. Mobile cone-beam computed tomography (CBCT) enables volumetric assessment of intraoperative anatomy beyond conventional point-based measurements. This study compared clinically applied distance measurement techniques with geometric distance descriptors retrospectively extracted from comprehensive three-dimensional CBCT datasets.
Methods
Ninety-six patients receiving IOeRT as a tumor-bed boost during breast-conserving surgery were included. Manual CBCT measurements (mCBCT) were compared with mean (USavg) and minimum (USmin) ultrasound measurements per patient. Comprehensive 3D distance extraction was performed in 82 CBCT datasets using an approximated thoracic-wall surface as the geometric reference. Minimum and first-quartile distances were evaluated across the complete applicator cross-section and within a 1-cm-diameter inner disk around the central beam axis.
Results
The mean paired difference between mCBCT and USavg was 0.3 mm and was not statistically significant, whereas mCBCT measurements exceeded USmin by an average of 1.5 mm. In the principal comparison, the minimum CBCT-derived distance within the inner disk was 1.4 mm smaller than mCBCT. Across the complete applicator area, minimum distances were 9.1-10.7 mm smaller than the clinical measurements. The first-quartile distance across all surface points was 1.4 mm smaller than mCBCT, whereas the corresponding inner-disk difference was not statistically significant after correction. Isodose-atlas-based dose-volume analysis indicated that a 1.5-mm change in assumed penetration depth could correspond to an approximately 7% change in the 90% isodose volume at shallow depths.
Conclusions
Clinical CBCT and ultrasound measurements showed small group-level differences but substantial patient-level variability. Comprehensive 3D CBCT analysis revealed spatial variations in tissue depth not fully represented by the clinical measurements, highlighting the limitations of characterizing treatment geometry using a single distance. Although anatomical accuracy and patient-specific dosimetric consequences remain to be established, these findings support the potential of intraoperative 3D CBCT to advance IOeRT toward individualized, image-guided treatment planning.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Paracelsus Medical University, 5020 Salzburg, Austria.
- Department of Gynecology, Paracelsus Medical University, 5020 Salzburg, Austria.
- radART-Institute for Research and Development on Advanced Radiation Technologies, 5020 Salzburg, Austria.
ORCID-Profile: Markus Stana, Sara Vockner, Ivan M Messner, Brane Grambozov, Johannes Berchtold, Philipp Scherer
Zitieren
APA
Stana, M., Ellmauer, K., Vockner, S., Messner, I. M., Fastner, G., Fussl, C., Grambozov, B., Kaiser, J., Berchtold, J., Sir, A., Scharle, N., Scherer, P., Gaisberger, C., Röder, F., & Mattke, M. (2026). A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy. Diagnostics, 16(21), 3664. https://doi.org/10.3390/diagnostics16213664
Vancouver
Stana M, Ellmauer K, Vockner S, Messner IM, Fastner G, Fussl C, et al. A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy. Diagnostics. 2026;16(21):3664. doi:10.3390/diagnostics16213664
BibTeX
@article{Stana2026,
author = {Stana, Markus and Ellmauer, Klarissa and Vockner, Sara and Messner, Ivan M and Fastner, Gerd and Fussl, Christoph and Grambozov, Brane and Kaiser, Julia and Berchtold, Johannes and Sir, Andreas and Scharle, Ninette and Scherer, Philipp and Gaisberger, Christoph and Röder, Falk and Mattke, Matthias},
title = {A 3D CBCT-Based Approach to Assessing Effective Radiation Penetration Depth in Intraoperative Electron Radiotherapy},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {21},
pages = {3664},
doi = {10.3390/diagnostics16213664}
}
EUVIMED