An Integrated Approach to Control the Penetration Depth of 3D-Printed Hollow Microneedles

Kendall Marie Defelippi, Allyson Yuuka Saumei Kwong, Julia Rose Appleget, Rana Altay, Maya Bree Matheny, Mary Margaret Dubus, Lily Marie Eribes, Maryam Mobed-Miremadi

Applied Mechanics · 22.03.2024 · 5 (2) · 233-259

DOI: 10.3390/applmech5020015 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Mechanics, online erschienen am 22.03.2024. Verfasst von Kendall Marie Defelippi und 7 weiteren Autor:innen. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 85 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science Foundation MRI Program, Santa Clara University’s School of Engineering Senior Design funds.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A variety of hollow microneedle (HMN) designs has emerged for minimally invasive therapies and monitoring systems. In this study, a design change limiting the indentation depth of the (3D) printed custom microneedle assembly (circular array of five conical frusta with and without a stopper, aspect ratio = 1.875) fabricated using stereolithography has been experimentally validated and modeled in silico. The micro-indentation profiles generated in confined compression on 1 mm ± 0.073 mm alginate films enabled the generation of a Prony series, where displacement ranged from 100 to 250 µm. These constants were used as intrinsic properties simulating experimental ramp/release profiles. Puncture occurred on two distinct hydrogel formulations at the design depth of 150 µm and indentation rate of 0.1 mm/s characterized by a peak force of 3.5 N (H = 31 kPa) and 8.3 N (H = 36.5 kPa), respectively. Experimental and theoretical alignments for peak force trends were obtained when the printing resolution was simulated. Higher puncture force and uniformity inferred by the stopper was confirmed via microscopy and profilometry. Meanwhile, poroviscoelasticity characterization is required to distinguish mass loss vs. redistribution post-indentation through pycnometry. Results from this paper highlight the feasibility of insertion-depth control within the epidermis thickness for the first time in solid HMN literature.

Förderung

  • National Science Foundation MRI Program
  • Santa Clara University’s School of Engineering Senior Design funds

Zitieren

APA

Defelippi, K. M., Kwong, A. Y. S., Appleget, J. R., Altay, R., Matheny, M. B., Dubus, M. M., Eribes, L. M., & Mobed-Miremadi, M. (2024). An Integrated Approach to Control the Penetration Depth of 3D-Printed Hollow Microneedles. Applied Mechanics, 5(2), 233-259. https://doi.org/10.3390/applmech5020015

Vancouver

Defelippi KM, Kwong AYS, Appleget JR, Altay R, Matheny MB, Dubus MM, et al. An Integrated Approach to Control the Penetration Depth of 3D-Printed Hollow Microneedles. Applied Mechanics. 2024;5(2):233-259. doi:10.3390/applmech5020015

BibTeX
@article{Defelippi2024,
  author = {Defelippi, Kendall Marie and Kwong, Allyson Yuuka Saumei and Appleget, Julia Rose and Altay, Rana and Matheny, Maya Bree and Dubus, Mary Margaret and Eribes, Lily Marie and Mobed-Miremadi, Maryam},
  title = {An Integrated Approach to Control the Penetration Depth of 3D-Printed Hollow Microneedles},
  journal = {Applied Mechanics},
  year = {2024},
  volume = {5},
  number = {2},
  pages = {233--259},
  doi = {10.3390/applmech5020015}
}
Kostenlos starten →