EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Effect of the Measurement Interface on Maximal Expiratory Pressure During Voluntary Cough in Young Adults

Rodrigo Muñoz-Cofré iD, Cristian Villar Gonzalez, Francisca Yáñez Quiñones, Paul Medina-González, Mariano Del Sol iD, Raúl Escobar Henríquez, Máximo Escobar-Cabello

Diagnostics · 03.11.2026 · 16 (21) · 3626

DOI: 10.3390/diagnostics16213626 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 03.11.2026. Verfasst von Rodrigo Muñoz-Cofré und 6 weiteren Autor:innen. Beteiligte Einrichtung: Postdoctorado en Ciencias Morfológicas, Universidad de La Frontera, Temuco 4811230, Chile. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: To determine whether there are differences between a nozzle and an oronasal mask in the measurement of maximal expiratory pressure (MEP) during a voluntary cough maneuver in young university students.

Methods

Forty university students aged 20–25 years were recruited, comprising 20 males and 20 females. Forced vital capacity and MEP were measured during a voluntary cough maneuver. All participants were assessed using both interfaces. The order of the interfaces used to measure MEP during the voluntary cough maneuver was randomized, once the order was determined, the evaluations were conducted on two separate days: one for the nozzle and another for the oronasal mask. Subsequently, participants were asked to report their perception of air leakage and comfort associated with each device.

Results

There was no significant main effect of the interface on MEP after adjusting for gender, peak expiratory flow (PEF), and body mass index (BMI) (F(1.36) = 3.285; p = 0.078; η2p = 0.084). There was a significant interaction between interface and BMI (F(1,36) = 6.391; p = 0.016; η2p = 0.151). Gender and PEF were significantly associated with MEP (gender: F(1,36) = 16.698; p < 0.001; η2p = 0.317; PEF: F(1,36) = 8.985; p = 0.005; η2p = 0.200).

Conclusions

In young adults, the measurement interface alone did not significantly alter MEP. The relationship between BMI and MEP varied according to the interface used, as evidenced by the interface × BMI interaction; gender and PEF were significantly associated with MEP.

Schlagwörter

Stichwörter der Autor:innen

coughmaximal espiratory pressurenozzleoronasal mask

Autor:innen und Institutionen

  • Postdoctorado en Ciencias Morfológicas, Universidad de La Frontera, Temuco 4811230, Chile.
  • Escuele de Kinesiología, Universidad Católica del Maule, Talca 4811230, Chile.
  • Centro de Excelencia en Estudios Morfológicos y Quirúrgicos, Universidad de La Frontera, Temuco 4811230, Chile.
  • Departamento de Neurología Pediátrica, Escuela de Medicina, Pontificia Universidad Católica de Chile, Santiago 8320165, Chile.

ORCID-Profile: Rodrigo Muñoz-Cofré, Mariano Del Sol

Zitieren

APA

Muñoz-Cofré, R., Gonzalez, C. V., Quiñones, F. Y., Medina-González, P., Sol, M. D., Henríquez, R. E., & Escobar-Cabello, M. (2026). Effect of the Measurement Interface on Maximal Expiratory Pressure During Voluntary Cough in Young Adults. Diagnostics, 16(21), 3626. https://doi.org/10.3390/diagnostics16213626

Vancouver

Muñoz-Cofré R, Gonzalez CV, Quiñones FY, Medina-González P, Sol MD, Henríquez RE, et al. Effect of the Measurement Interface on Maximal Expiratory Pressure During Voluntary Cough in Young Adults. Diagnostics. 2026;16(21):3626. doi:10.3390/diagnostics16213626

BibTeX
@article{MuozCofr2026,
  author = {Muñoz-Cofré, Rodrigo and Gonzalez, Cristian Villar and Quiñones, Francisca Yáñez and Medina-González, Paul and Sol, Mariano Del and Henríquez, Raúl Escobar and Escobar-Cabello, Máximo},
  title = {Effect of the Measurement Interface on Maximal Expiratory Pressure During Voluntary Cough in Young Adults},
  journal = {Diagnostics},
  year = {2026},
  volume = {16},
  number = {21},
  pages = {3626},
  doi = {10.3390/diagnostics16213626}
}
Kostenlos starten →