Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study
Diagnostics · 20.08.2026 · 16 (16) · 2652
DOI: 10.3390/diagnostics16162652 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 20.08.2026. Verfasst von Donghyo Yang und 4 weiteren Autor:innen. Beteiligte Einrichtung: College of Dentistry, Chosun University, Gwangju 61452, Republic of Korea. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Chosun University Dental Hospital.
Abstract (Originalsprache)
Background/Objectives: Cone-beam computed tomography (CBCT)-to-intraoral-scan registration in edentulous implant planning relies on radiopaque fiducial markers, yet how marker design governs marker-geometry reproduction is unclear. We quantified the effects of shape, size and configuration and where the error arose.
Methods
Eighteen edentulous mandibular denture designs (three shapes × three sizes, 2–4 mm × two configurations; five markers each) were printed and CBCT-imaged three times at 0.2 mm voxel. Distances, angles, a leave-one-marker-out (LOMO) error and a markerless control were analyzed at the design level (n = 18).
Results
Acquisition reproducibility was voxel-limited (median 49 μm). Marker size governed trueness and precision (partial ω2p 0.25–0.65), with no shape effect detected. The LOMO error fell with size (541 to 359 μm; p = 0.004), whereas the markerless control did not (p = 0.653); the size effect differed between approaches. Twenty percent of 2 mm markers exceeded the 300 μm threshold—unstable localization, not failed detection. In the markerless control, 6 of 270 markers (2.2%) exceeded 1 mm despite acceptable global surface fits.
Conclusions
A satisfactory global surface-fit residual may conceal substantial local misregistration, whereas fiducial landmarks permit independent verification of alignment. Marker size most directly governed reproduction fidelity; ≥3 mm diameters are recommended for the rotationally symmetric geometries tested.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- College of Dentistry, Chosun University, Gwangju 61452, Republic of Korea.
ORCID-Profile: Joo-Hun Song
Förderung
- Chosun University Dental Hospital
Zitieren
APA
Yang, D., Kim, G., Hwang, S., Seo, Y. S., & Song, J. H. (2026). Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study. Diagnostics, 16(16), 2652. https://doi.org/10.3390/diagnostics16162652
Vancouver
Yang D, Kim G, Hwang S, Seo YS, Song JH. Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study. Diagnostics. 2026;16(16):2652. doi:10.3390/diagnostics16162652
BibTeX
@article{Yang2026,
author = {Yang, Donghyo and Kim, Gahyun and Hwang, Seoyun and Seo, Yo-Seob and Song, Joo-Hun},
title = {Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {16},
pages = {2652},
doi = {10.3390/diagnostics16162652}
}
EUVIMED