Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study

Donghyo Yang, Gahyun Kim, Seoyun Hwang, Yo-Seob Seo, Joo-Hun Song iD

Diagnostics · 20.08.2026 · 16 (16) · 2652

DOI: 10.3390/diagnostics16162652 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 20.08.2026. Verfasst von Donghyo Yang und 4 weiteren Autor:innen. Beteiligte Einrichtung: College of Dentistry, Chosun University, Gwangju 61452, Republic of Korea. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Chosun University Dental Hospital.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Cone-beam computed tomography (CBCT)-to-intraoral-scan registration in edentulous implant planning relies on radiopaque fiducial markers, yet how marker design governs marker-geometry reproduction is unclear. We quantified the effects of shape, size and configuration and where the error arose.

Methods

Eighteen edentulous mandibular denture designs (three shapes × three sizes, 2–4 mm × two configurations; five markers each) were printed and CBCT-imaged three times at 0.2 mm voxel. Distances, angles, a leave-one-marker-out (LOMO) error and a markerless control were analyzed at the design level (n = 18).

Results

Acquisition reproducibility was voxel-limited (median 49 μm). Marker size governed trueness and precision (partial ω2p 0.25–0.65), with no shape effect detected. The LOMO error fell with size (541 to 359 μm; p = 0.004), whereas the markerless control did not (p = 0.653); the size effect differed between approaches. Twenty percent of 2 mm markers exceeded the 300 μm threshold—unstable localization, not failed detection. In the markerless control, 6 of 270 markers (2.2%) exceeded 1 mm despite acceptable global surface fits.

Conclusions

A satisfactory global surface-fit residual may conceal substantial local misregistration, whereas fiducial landmarks permit independent verification of alignment. Marker size most directly governed reproduction fidelity; ≥3 mm diameters are recommended for the rotationally symmetric geometries tested.

Schlagwörter

Stichwörter der Autor:innen

cone-beam computed tomographydental implantsedentulous jawfiducial markersimage registrationintraoral scanningprecisiontrueness

Autor:innen und Institutionen

  • College of Dentistry, Chosun University, Gwangju 61452, Republic of Korea.

ORCID-Profile: Joo-Hun Song

Förderung

  • Chosun University Dental Hospital

Zitieren

APA

Yang, D., Kim, G., Hwang, S., Seo, Y. S., & Song, J. H. (2026). Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study. Diagnostics, 16(16), 2652. https://doi.org/10.3390/diagnostics16162652

Vancouver

Yang D, Kim G, Hwang S, Seo YS, Song JH. Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study. Diagnostics. 2026;16(16):2652. doi:10.3390/diagnostics16162652

BibTeX
@article{Yang2026,
  author = {Yang, Donghyo and Kim, Gahyun and Hwang, Seoyun and Seo, Yo-Seob and Song, Joo-Hun},
  title = {Effect of Fiducial Marker Design (Shape, Size, and Configuration) on the Trueness and Precision of CBCT Marker-Geometry Reproduction for CBCT–Intraoral Scan Registration: An In Vitro Study},
  journal = {Diagnostics},
  year = {2026},
  volume = {16},
  number = {16},
  pages = {2652},
  doi = {10.3390/diagnostics16162652}
}
Kostenlos starten →