EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery

Se An Oh iD, Sung Yeop Kim, Jae Won Park, Ji Woon Yea, Jaehyeon Park iD, Yoon Young Jo iD

Diagnostics · 19.08.2026 · 16 (16) · 2628

DOI: 10.3390/diagnostics16162628 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 19.08.2026. Verfasst von Se An Oh und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Yeungnam University Medical Center, Daegu 42415, Republic of Korea. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Yeungnam University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Accurate verification of single-isocenter multiple-target (SIMT) stereotactic radiosurgery is challenging owing to the complexity of multi-lesion delivery and the sensitivity of stereotactic dose gradients. We aimed to evaluate the efficacy of electronic portal imaging device (EPID)-based multileaf collimator (MLC) transmission and dosimetric leaf gap (DLG) correction for improving portal-dose prediction agreement in SIMT stereotactic radiosurgery (SRS) and to propose an exploratory target count-based institutional action level.

Methods

This retrospective analysis included 112 consecutive patients treated with HyperArc™-based SRS (1–13 targets). Treatment plans were calculated using the Acuros XB algorithm for 6 MV flattening filter-free (FFF) beams. Two sets of MLC parameters were compared for portal-dose image prediction (PDIP): (1) standard parameters measured using an ion chamber; (2) EPID-derived corrected parameters. Portal-dose accuracy was evaluated using gamma index analysis with a 95% pass rate threshold.

Results

The uncorrected method showed a strong negative correlation between target number and gamma passing rates, with complex plans (≥9 targets) dropping as low as 70%. EPID-based correction substantially improved dose agreement, yielding consistent passing rates above 98%, regardless of target number. Although uncorrected parameters remained within tolerance for plans with one to two targets, accuracy declined markedly starting at three targets.

Conclusions

Our findings indicate that the EPID-based correction of MLC transmission and DLG mitigated cumulative modeling discrepancies in complex SIMT SRS. Thus, we propose the exploratory institutional observation of three targets, beyond which EPID-based correction may be considered to achieve optimal portal-dose prediction agreement in HyperArc-based stereotactic radiosurgery.

Schlagwörter

Stichwörter der Autor:innen

HyperArcdosimetric leaf gap (DLG)electronic portal imaging device (EPID)portal dosimetryradiosurgerysingle-isocenter multiple targets (SIMT)

Autor:innen und Institutionen

  • Department of Radiation Oncology, Yeungnam University Medical Center, Daegu 42415, Republic of Korea.
  • Department of Radiation Oncology, College of Medicine, Yeungnam University, Daegu 42415, Republic of Korea.

ORCID-Profile: Se An Oh, Jaehyeon Park, Yoon Young Jo

Förderung

  • Yeungnam University

Zitieren

APA

Oh, S. A., Kim, S. Y., Park, J. W., Yea, J. W., Park, J., & Jo, Y. Y. (2026). Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery. Diagnostics, 16(16), 2628. https://doi.org/10.3390/diagnostics16162628

Vancouver

Oh SA, Kim SY, Park JW, Yea JW, Park J, Jo YY. Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery. Diagnostics. 2026;16(16):2628. doi:10.3390/diagnostics16162628

BibTeX
@article{Oh2026,
  author = {Oh, Se An and Kim, Sung Yeop and Park, Jae Won and Yea, Ji Woon and Park, Jaehyeon and Jo, Yoon Young},
  title = {Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery},
  journal = {Diagnostics},
  year = {2026},
  volume = {16},
  number = {16},
  pages = {2628},
  doi = {10.3390/diagnostics16162628}
}
Kostenlos starten →