Layer-Specific Shear Wave Elastography Reveals the Tissue Origin of MyotonPRO-Derived Muscle Stiffness
Diagnostics · 27.07.2026 · 16 (15) · 2346
DOI: 10.3390/diagnostics16152346 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 27.07.2026. Verfasst von Sebastian Szajkowski und 3 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medical Sciences, Warsaw Medical Academy of Applied Sciences, 8 Rydygiera St., 01-793 Warszawa, Poland. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Shear wave elastography (SWE) and myotonometry are commonly used to assess muscle stiffness, but they reflect different aspects of tissue mechanics. SWE allows layer-specific evaluation, whereas myotonometry captures the response of a broader tissue complex. The contribution of individual layers to MyotonPRO-derived stiffness remains unclear. This study examined the relationship between MyotonPRO stiffness and layer-specific SWE measurements in the tibialis anterior muscle.
Methods
In this cross-sectional study, 56 healthy adults were evaluated. Muscle stiffness was assessed using MyotonPRO and SWE across three regions of interest: superficial tissues including skin, subcutaneous tissue, and deep fascia (ROI 1); muscle tissue excluding fascia (ROI 2); and a composite region including all tissues from the skin surface to a depth of 2 cm (ROI 3). Relationships were analyzed using Spearman correlation, multiple linear regression, and Bland–Altman analysis.
Results
Moderate but consistent correlations were observed between MyotonPRO stiffness and SWE values in ROI 2 (ρ = 0.516, p < 0.001) and ROI 3 (ρ = 0.550, p < 0.001), whereas no significant association was found in ROI 1. SWE values in ROI 2 and ROI 3 were significant predictors of MyotonPRO stiffness, explaining 38% of the variance. Bland–Altman analysis demonstrated moderate agreement with a systematic bias toward higher MyotonPRO values.
Conclusions
MyotonPRO-derived stiffness reflects a composite mechanical response of muscle and surrounding tissues, with the strongest associations observed in regions encompassing both muscle and fascia. These findings highlight the importance of tissue composition and ROI selection in stiffness assessment and may facilitate more accurate interpretation of MyotonPRO measurements in musculoskeletal diagnostics and rehabilitation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Medical Sciences, Warsaw Medical Academy of Applied Sciences, 8 Rydygiera St., 01-793 Warszawa, Poland.
- Medical Rehabilitation and Osteopathy Clinic REHApunkt, Traugutta 12 St., 05-820 Piastów, Poland.
- Department of Sports Medicine and Physiology of Physical Exercise, Faculty of Health Science in Katowice, Medical University of Silesia, 40-055 Katowice, Poland.
- Collegium Medicum im dr Władysława Biegańskiego, Jan Długosz University in Częstochowa, 13/15 Armii Krajowej St., 42-200 Częstochowa, Poland.
ORCID-Profile: Sebastian Szajkowski, Michał Dwornik, Krzysztof Suszyński, Jarosław Pasek
Zitieren
APA
Szajkowski, S., Dwornik, M., Suszyński, K., & Pasek, J. (2026). Layer-Specific Shear Wave Elastography Reveals the Tissue Origin of MyotonPRO-Derived Muscle Stiffness. Diagnostics, 16(15), 2346. https://doi.org/10.3390/diagnostics16152346
Vancouver
Szajkowski S, Dwornik M, Suszyński K, Pasek J. Layer-Specific Shear Wave Elastography Reveals the Tissue Origin of MyotonPRO-Derived Muscle Stiffness. Diagnostics. 2026;16(15):2346. doi:10.3390/diagnostics16152346
BibTeX
@article{Szajkowski2026,
author = {Szajkowski, Sebastian and Dwornik, Michał and Suszyński, Krzysztof and Pasek, Jarosław},
title = {Layer-Specific Shear Wave Elastography Reveals the Tissue Origin of MyotonPRO-Derived Muscle Stiffness},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {15},
pages = {2346},
doi = {10.3390/diagnostics16152346}
}
EUVIMED