The Impact of Automatic Exposure Control Technology on the In Vivo Radiation Dose in Digital Mammography: A Comparison Between Different Systems and Target/Filter Combinations

Ahmad A Alhulail iD, Salman M Albeshan iD, Mohammed S Alshuhri iD, Essam M Alkhybari, Mansour A Almanaa iD, Haitham Alahmad iD, Khaled Alenazi iD, Abdulaziz S Alshabibi, Mohammed Alsufayan, Saleh A Alsulaiman, Maha M Almuqbil, Mahmoud M Elsharkawi iD und 1 weitere

Diagnostics · 08.05.2025 · 15 (10) · 1185

DOI: 10.3390/diagnostics15101185 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 08.05.2025. Verfasst von Ahmad A Alhulail und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Medical Imaging, Prince Sattam bin Abdulaziz University, Al-Kharj 16278, Saudi Arabia. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Digital mammography is widely used for breast cancer screening; however, variations in system design and automatic exposure control (AEC) strategies can lead to significant differences in radiation dose, potentially affecting the diagnostic quality and patient safety. In this study, we aimed to determine the effect of various mammographic technologies on the in vivo mean glandular doses (MGDs) that are received in clinical settings.

Methods

The MGDs and applied acquisition parameters from 194,608 mammograms, acquired employing AEC using different digital mammography systems (GE, Siemens, and two different models of Hologic), were retrospectively collected. The potential variation in MGD resulting from different technologies (system and target/filter combination) was assessed employing the Kruskal–Wallis test, followed by Dunn’s post hoc. The AEC optimization of acquisition parameters (kVp, mAs) within each system was investigated through a multi-regression analysis as a function of the compressed breast thickness (CBT). The trend line of these parameters in addition to the MGD and source-to-breast distance were also plotted and compared.

Results

There were significant variations in delivered doses per CBT based on which technology was used (p < 0.001). The regression analyses revealed system-specific differences in AEC adjustments of mAs and kVp in response to CBT changes. As the CBT increases, the MGD increases with different degrees, rates, and patterns across systems due to differences in AEC strategies.

Conclusions

The MGD is affected by the applied technology, which is different between systems. Clinicians need to be aware of these variations and how they affect the MGD. Additionally, manufacturers may need to consider standardizing the implemented technology effects on the MGDs.

Schlagwörter

Stichwörter der Autor:innen

AEC optimizationcompressed breast thicknessdigital mammographymammography technologymean glandular dosetarget/filter combination

Autor:innen und Institutionen

  • Department of Radiology and Medical Imaging, Prince Sattam bin Abdulaziz University, Al-Kharj 16278, Saudi Arabia.
  • Department of Radiological Sciences, College of Applied Medical Sciences, King Saud University, Riyadh 11451, Saudi Arabia.
  • Radiology Department, Altakassusi Alliance Medical LLC, King Fahad Medical City, Riyadh 12231, Saudi Arabia.
  • Radiology & Medical Imaging Department, Imam Abdulrahman Alfisal Hospital, Cluster One, Riyadh 14723, Saudi Arabia.
  • Cancer Control Program, Ministry of Health, Riyadh 12628, Saudi Arabia.
  • Department of Health Programs, Ministry of Health, Riyadh 12628, Saudi Arabia.

ORCID-Profile: Ahmad A Alhulail, Salman M Albeshan, Mohammed S Alshuhri, Mansour A Almanaa, Haitham Alahmad, Khaled Alenazi, Mahmoud M Elsharkawi

Zitieren

APA

Alhulail, A. A., Albeshan, S. M., Alshuhri, M. S., Alkhybari, E. M., Almanaa, M. A., Alahmad, H., Alenazi, K., Alshabibi, A. S., Alsufayan, M., Alsulaiman, S. A., Almuqbil, M. M., Elsharkawi, M. M., & Alghamdi, S. (2025). The Impact of Automatic Exposure Control Technology on the In Vivo Radiation Dose in Digital Mammography: A Comparison Between Different Systems and Target/Filter Combinations. Diagnostics, 15(10), 1185. https://doi.org/10.3390/diagnostics15101185

Vancouver

Alhulail AA, Albeshan SM, Alshuhri MS, Alkhybari EM, Almanaa MA, Alahmad H, et al. The Impact of Automatic Exposure Control Technology on the In Vivo Radiation Dose in Digital Mammography: A Comparison Between Different Systems and Target/Filter Combinations. Diagnostics. 2025;15(10):1185. doi:10.3390/diagnostics15101185

BibTeX
@article{Alhulail2025,
  author = {Alhulail, Ahmad A and Albeshan, Salman M and Alshuhri, Mohammed S and Alkhybari, Essam M and Almanaa, Mansour A and Alahmad, Haitham and Alenazi, Khaled and Alshabibi, Abdulaziz S and Alsufayan, Mohammed and Alsulaiman, Saleh A and Almuqbil, Maha M and Elsharkawi, Mahmoud M and Alghamdi, Sultan},
  title = {The Impact of Automatic Exposure Control Technology on the In Vivo Radiation Dose in Digital Mammography: A Comparison Between Different Systems and Target/Filter Combinations},
  journal = {Diagnostics},
  year = {2025},
  volume = {15},
  number = {10},
  pages = {1185},
  doi = {10.3390/diagnostics15101185}
}
Kostenlos starten →