Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes
Diagnostics · 18.12.2023 · 13 (24) · 3687
DOI: 10.3390/diagnostics13243687 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 18.12.2023. Verfasst von Kyung-Sun Na und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Yeouido St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 16247, Republic of Korea. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fondazione Roma.
Abstract (Originalsprache)
Purpose
To analyse the correspondence between refractive astigmatism and corneal astigmatism in pseudophakic eyes with non-toric intraocular lenses.
Setting
Yeouido St. Mary hospital, Seoul, Republic of Korea.
Design
Evaluation of a diagnostic test instrument.
Methods
This retrospective study included 95 eyes of 95 patients. Corneal astigmatism was measured with an automated keratometer (RK-5, Canon) and Scheimpflug tomography (Pentacam HR, Oculus). Refractive astigmatism was compared to keratometric astigmatism (based on anterior corneal measurements only), equivalent K-reading, and total corneal astigmatism (both based on anterior and posterior corneal measurements). Vector analysis was carried out by Næser’s polar value method. The accuracy was defined as the average magnitude of the vectorial difference in astigmatism (DA). Each corneal measurement was optimized in retrospect by a multiple linear regression equation between refractive and corneal astigmatism.
Results
Keratometric astigmatism overestimated with-the-rule (WTR) refractive astigmatism and underestimated against-the-rule (ATR) refractive astigmatism. Several measurements based on both corneal surfaces’ values did not show any statistically significant difference with respect to refractive astigmatism. The mean corneal astigmatism by total corneal refractive power (TCRP) at 4.0 mm (zone/pupil) produced the lowest mean arithmetic DA and the highest percentage of eyes with a DA ≤ 0.50 dioptre. After optimization, the accuracies of automated KA and TCRP 4.0 mm (zone/pupil) were similar.
Conclusions
Total corneal astigmatism measured by Scheimpflug tomography at a 4.0 mm zone centered on the pupil accurately reflects the refractive astigmatism in pseudophakic eyes. However, the accuracy of total corneal astigmatism is not different from automated KA after optimization.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology, Yeouido St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 16247, Republic of Korea.
- IRCCS-G.B. Bietti Foundation, 00184 Rome, Italy.
- Randers Regional Hospital, 8930 Randers, Denmark.
Förderung
- Fondazione Roma
Zitieren
APA
Na, K. S., Savini, G., Whang, W. J., & Næser, K. (2023). Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes. Diagnostics, 13(24), 3687. https://doi.org/10.3390/diagnostics13243687
Vancouver
Na KS, Savini G, Whang WJ, Næser K. Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes. Diagnostics. 2023;13(24):3687. doi:10.3390/diagnostics13243687
BibTeX
@article{Na2023,
author = {Na, Kyung-Sun and Savini, Giacomo and Whang, Woong-Joo and Næser, Kristian},
title = {Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes},
journal = {Diagnostics},
year = {2023},
volume = {13},
number = {24},
pages = {3687},
doi = {10.3390/diagnostics13243687}
}
EUVIMED