Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes

Kyung-Sun Na, Giacomo Savini, Woong-Joo Whang, Kristian Næser

Diagnostics · 18.12.2023 · 13 (24) · 3687

DOI: 10.3390/diagnostics13243687 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 18.12.2023. Verfasst von Kyung-Sun Na und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Yeouido St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 16247, Republic of Korea. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fondazione Roma.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To analyse the correspondence between refractive astigmatism and corneal astigmatism in pseudophakic eyes with non-toric intraocular lenses.

Setting

Yeouido St. Mary hospital, Seoul, Republic of Korea.

Design

Evaluation of a diagnostic test instrument.

Methods

This retrospective study included 95 eyes of 95 patients. Corneal astigmatism was measured with an automated keratometer (RK-5, Canon) and Scheimpflug tomography (Pentacam HR, Oculus). Refractive astigmatism was compared to keratometric astigmatism (based on anterior corneal measurements only), equivalent K-reading, and total corneal astigmatism (both based on anterior and posterior corneal measurements). Vector analysis was carried out by Næser’s polar value method. The accuracy was defined as the average magnitude of the vectorial difference in astigmatism (DA). Each corneal measurement was optimized in retrospect by a multiple linear regression equation between refractive and corneal astigmatism.

Results

Keratometric astigmatism overestimated with-the-rule (WTR) refractive astigmatism and underestimated against-the-rule (ATR) refractive astigmatism. Several measurements based on both corneal surfaces’ values did not show any statistically significant difference with respect to refractive astigmatism. The mean corneal astigmatism by total corneal refractive power (TCRP) at 4.0 mm (zone/pupil) produced the lowest mean arithmetic DA and the highest percentage of eyes with a DA ≤ 0.50 dioptre. After optimization, the accuracies of automated KA and TCRP 4.0 mm (zone/pupil) were similar.

Conclusions

Total corneal astigmatism measured by Scheimpflug tomography at a 4.0 mm zone centered on the pupil accurately reflects the refractive astigmatism in pseudophakic eyes. However, the accuracy of total corneal astigmatism is not different from automated KA after optimization.

Schlagwörter

Stichwörter der Autor:innen

Scheimpflug tomographyastigmatismcorneal curvature

Autor:innen und Institutionen

  • Department of Ophthalmology, Yeouido St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul 16247, Republic of Korea.
  • IRCCS-G.B. Bietti Foundation, 00184 Rome, Italy.
  • Randers Regional Hospital, 8930 Randers, Denmark.

Förderung

  • Fondazione Roma

Zitieren

APA

Na, K. S., Savini, G., Whang, W. J., & Næser, K. (2023). Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes. Diagnostics, 13(24), 3687. https://doi.org/10.3390/diagnostics13243687

Vancouver

Na KS, Savini G, Whang WJ, Næser K. Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes. Diagnostics. 2023;13(24):3687. doi:10.3390/diagnostics13243687

BibTeX
@article{Na2023,
  author = {Na, Kyung-Sun and Savini, Giacomo and Whang, Woong-Joo and Næser, Kristian},
  title = {Comparison of Automated Keratometer and Scheimpflug Tomography for Predicting Refractive Astigmatism in Pseudophakic Eyes},
  journal = {Diagnostics},
  year = {2023},
  volume = {13},
  number = {24},
  pages = {3687},
  doi = {10.3390/diagnostics13243687}
}
Kostenlos starten →