An Intra-Individual Comparison of Low-keV Photon-Counting CT versus Energy-Integrating-Detector CT Angiography of the Aorta
Diagnostics · 12.12.2023 · 13 (24) · 3645
DOI: 10.3390/diagnostics13243645 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 12.12.2023. Verfasst von Jan-Lucca Hennes und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, 97080 Würzburg, Germany. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Interdisciplinary Center of Clinical Research Würzburg, Deutsche Forschungsgemeinschaft, Open Access Publication Fund of the University of Würzburg.
Abstract (Originalsprache)
This retrospective study aims to provide an intra-individual comparison of aortic CT angiographies (CTAs) using first-generation photon-counting-detector CT (PCD-CT) and third-generation energy-integrating-detector CT (EID-CT). High-pitch CTAs were performed with both scanners and equal contrast-agent protocols. EID-CT employed automatic tube voltage selection (90/100 kVp) with reference tube current of 434/350 mAs, whereas multi-energy PCD-CT scans were generated with fixed tube voltage (120 kVp), image quality level of 64, and reconstructed as 55 keV monoenergetic images. For image quality assessment, contrast-to-noise ratios (CNRs) were calculated, and subjective evaluation (overall quality, luminal contrast, vessel sharpness, blooming, and beam hardening) was performed independently by three radiologists. Fifty-seven patients (12 women, 45 men) were included with a median interval between examinations of 12.7 months (interquartile range 11.1 months). Using manufacturer-recommended scan protocols resulted in a substantially lower radiation dose in PCD-CT (size-specific dose estimate: 4.88 ± 0.48 versus 6.28 ± 0.50 mGy, p < 0.001), while CNR was approximately 50% higher (41.11 ± 8.68 versus 27.05 ± 6.73, p < 0.001). Overall image quality and luminal contrast were deemed superior in PCD-CT (p < 0.001). Notably, EID-CT allowed for comparable vessel sharpness (p = 0.439) and less pronounced blooming and beam hardening (p < 0.001). Inter-rater agreement was good to excellent (0.58–0.87). Concluding, aortic PCD-CTAs facilitate increased image quality with significantly lower radiation dose compared to EID-CTAs.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, 97080 Würzburg, Germany.
ORCID-Profile: Henner Huflage, Jan-Peter Grunz, Viktor Hartung, Theresa Sophie Patzer, Philipp Gruschwitz
Förderung
- Interdisciplinary Center of Clinical Research Würzburg
- Deutsche Forschungsgemeinschaft
- Open Access Publication Fund of the University of Würzburg
- Department of Diagnostic and Interventional Radiology at University Hospital Würzburg receives research
Zitieren
APA
Hennes, J. L., Huflage, H., Grunz, J. P., Hartung, V., Augustin, A. M., Patzer, T. S., Pannenbecker, P., Petritsch, B., Bley, T. A., & Gruschwitz, P. (2023). An Intra-Individual Comparison of Low-keV Photon-Counting CT versus Energy-Integrating-Detector CT Angiography of the Aorta. Diagnostics, 13(24), 3645. https://doi.org/10.3390/diagnostics13243645
Vancouver
Hennes JL, Huflage H, Grunz JP, Hartung V, Augustin AM, Patzer TS, et al. An Intra-Individual Comparison of Low-keV Photon-Counting CT versus Energy-Integrating-Detector CT Angiography of the Aorta. Diagnostics. 2023;13(24):3645. doi:10.3390/diagnostics13243645
BibTeX
@article{Hennes2023,
author = {Hennes, Jan-Lucca and Huflage, Henner and Grunz, Jan-Peter and Hartung, Viktor and Augustin, Anne Marie and Patzer, Theresa Sophie and Pannenbecker, Pauline and Petritsch, Bernhard and Bley, Thorsten Alexander and Gruschwitz, Philipp},
title = {An Intra-Individual Comparison of Low-keV Photon-Counting CT versus Energy-Integrating-Detector CT Angiography of the Aorta},
journal = {Diagnostics},
year = {2023},
volume = {13},
number = {24},
pages = {3645},
doi = {10.3390/diagnostics13243645}
}
EUVIMED