Automated coronary analysis in ultrahigh-spatial resolution photon-counting detector CT angiography: Clinical validation and intra-individual comparison with energy-integrating detector CT
Journal of Cardiovascular Computed Tomography · 2025-09 · 19 (5) · 576-584
DOI: 10.1016/j.jcct.2025.08.006 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Computed Tomography, 2025-09. Verfasst von Dmitrij Kravchenko und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Radiological Science, Medical University of South Carolina, SC, USA; Department of Diagnostic and Interventional Radiolog…. Hauptthemen laut MEDLINE: Computed Tomography Angiography, Coronary Angiography, Coronary Artery Disease, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Siemens Healthineers.
Abstract (Originalsprache)
OBJECTIVES: To evaluate a deep-learning algorithm for automated coronary artery analysis on ultrahigh-resolution photon-counting detector coronary computed tomography (CT) angiography and compared its performance to expert readers using invasive coronary angiography as reference. METHODS: Thirty-two patients (mean age 68.6 years; 81 % male) underwent both energy-integrating detector and ultrahigh-resolution photon-counting detector CT within 30 days. Expert readers scored each image using the Coronary Artery Disease-Reporting and Data System classification, and compared to invasive angiography. After a three-month wash-out, one reader reanalyzed the photon-counting detector CT images assisted by the algorithm. Sensitivity, specificity, accuracy, inter-reader agreement, and reading times were recorded for each method. RESULTS: On 401 arterial segments, inter-reader agreement improved from substantial (κ = 0.75) on energy-integrating detector CT to near-perfect (κ = 0.86) on photon-counting detector CT. The algorithm alone achieved 85 % sensitivity, 91 % specificity, and 90 % accuracy on energy-integrating detector CT, and 85 %, 96 %, and 95 % on photon-counting detector CT. Compared to invasive angiography on photon-counting detector CT, manual and automated reads had similar sensitivity (67 %), but manual assessment slightly outperformed regarding specificity (85 % vs. 79 %) and accuracy (84 % vs. 78 %). When the reader was assisted by the algorithm, specificity rose to 97 % (p < 0.001), accuracy to 95 %, and reading time decreased by 54 % (p < 0.001). CONCLUSION: This deep-learning algorithm demonstrates high agreement with experts and improved diagnostic performance on photon-counting detector CT. Expert review augmented by the algorithm further increases specificity and dramatically reduces interpretation time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology and Radiological Science, Medical University of South Carolina, SC, USA; Department of Diagnostic and Interventional Radiology, University Hospital Bonn, Bonn, Germany; Quantitative Imaging Labor…
- Department of Radiology and Radiological Science, Medical University of South Carolina, SC, USA; Department of Diagnostic and Interventional Radiology, Medical Center - University of Freiburg, Freiburg, Germany.
- Department of Radiology and Radiological Science, Medical University of South Carolina, SC, USA. Electronic address: musccvi@musc.edu.
- Department of Radiology and Radiological Science, Medical University of South Carolina, SC, USA.
- Computed Tomography, Siemens Healthineers, Forchheim, Germany.
- Department of Radiology and Radiological Science, Medical University of South Carolina, SC, USA; Siemens Medical Solutions, Malvern, PA, USA.
Förderung
- Siemens Healthineers
Zitieren
APA
Kravchenko, D., Hagar, M. T., Varga-Szemes, A., Schoepf, U. J., Schoebinger, M., O'Doherty, J., Gülsün, M. A., Laghi, A., Laux, G. S., Vecsey-Nagy, M., Emrich, T., & Tremamunno, G. (2025). Automated coronary analysis in ultrahigh-spatial resolution photon-counting detector CT angiography: Clinical validation and intra-individual comparison with energy-integrating detector CT. Journal of Cardiovascular Computed Tomography, 19(5), 576-584. https://doi.org/10.1016/j.jcct.2025.08.006
Vancouver
Kravchenko D, Hagar MT, Varga-Szemes A, Schoepf UJ, Schoebinger M, O'Doherty J, et al. Automated coronary analysis in ultrahigh-spatial resolution photon-counting detector CT angiography: Clinical validation and intra-individual comparison with energy-integrating detector CT. Journal of Cardiovascular Computed Tomography. 2025;19(5):576-584. doi:10.1016/j.jcct.2025.08.006
BibTeX
@article{Kravchenko2025,
author = {Kravchenko, Dmitrij and Hagar, Muhammad Taha and Varga-Szemes, Akos and Schoepf, U Joseph and Schoebinger, Max and O'Doherty, Jim and Gülsün, Mehmet Akif and Laghi, Andrea and Laux, Gerald Siegfried and Vecsey-Nagy, Milan and Emrich, Tilman and Tremamunno, Giuseppe},
title = {Automated coronary analysis in ultrahigh-spatial resolution photon-counting detector CT angiography: Clinical validation and intra-individual comparison with energy-integrating detector CT},
journal = {Journal of Cardiovascular Computed Tomography},
year = {2025},
volume = {19},
number = {5},
pages = {576--584},
doi = {10.1016/j.jcct.2025.08.006}
}
EUVIMED