Evaluation of the Versius Robotic System for Infant Surgery—A Study in Piglets of Less than 10 kg Body Weight
Children · 03.05.2023 · 10 (5) · 831
DOI: 10.3390/children10050831 ↗Kurz gefasst
Fachartikel in der Zeitschrift Children, online erschienen am 03.05.2023. Verfasst von Thomas Franz Krebs und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of General-, Visceral-, Thoracic-, Transplant- and Pediatric Surgery, UKSH University Hospital of Schleswig-Holstein Kiel Campus, Arnold-H…. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: CSRD VA, CMR Surgical (CMR Surgical Ltd, 194 Science Park, Milton Road, Cambridge, UK).
Abstract (Originalsprache)
Background
We were able to demonstrate the feasibility of a new robotic system (Versius, CMR Surgical, Cambridge, UK) for procedures in small inanimate cavities. The aim of this consecutive study was to test the Versius® system for its feasibility, performance, and safety of robotic abdominal and thoracic surgery in piglets simulating infants with a body weight lower than 10 kg.
Methods
A total of 24 procedures (from explorative laparoscopy to thoracoscopic esophageal repair) were performed in 4 piglets with a mean age of 12 days and a mean body weight of 6.4 (7–7.5) kg. Additional urological procedures were performed after euthanasia of the piglet. The Versius® robotic system was used with 5 mm wristed instruments and a 10 mm 3D 0° or 30° camera. The setup consisted of the master console and three to four separate arms. The performance of the procedure, the size, position, and the distance between the ports, the external and internal collisions, and complications of the procedures were recorded and analyzed.
Results
We were able to perform all surgical procedures as planned. We encountered neither surgical nor robot-associated complications in the live model. Whereas all abdominal procedures could be performed successfully under general anesthesia, one piglet was euthanized early before the thoracic interventions, likely due to pulmonary inflammatory response. Technical limitations were based on the size of the camera (10 mm) being too large and the minimal insertion depth of the instruments for calibration of the fulcrum point.
Conclusions
Robotic surgery on newborns and infants appears technically feasible with the Versius® system. Software adjustments for fulcrum point calibration need to be implemented by the manufacturer as a result of our study. To further evaluate the Versius® system, prospective trials are needed, comparing it to open and laparoscopic surgery as well as to other robotic systems.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of General-, Visceral-, Thoracic-, Transplant- and Pediatric Surgery, UKSH University Hospital of Schleswig-Holstein Kiel Campus, Arnold-Heller-Strasse 3, 24105 Kiel, Germany.
- Department of Pediatric Surgery, Ostschweizer Children's Hospital, Claudiusstrasse 6, 9006 St. Gallen, Switzerland.
- Department of Anesthesia and Intensive Care Medicine, UKSH University Hospital of Schleswig-Holstein Kiel Campus, Arnold-Heller-Strasse 3, 24105 Kiel, Germany.
- Department of Anesthesia and Intensive Care Medicine, Ev. Amalie Sieveking Hospital, Haselkamp 33, 22359 Hamburg, Germany.
- Department of Paediatric Surgery and Urology, Southampton Children's Hospital, University Hospital Southampton, Southampton SO16 6YD, UK.
- Kurt-Semm-Center for Minimally Invasive and Robotic Surgery, UKSH University Hospital of Schleswig-Holstein Kiel Campus, Arnold-Heller-Strasse 3, 24105 Kiel, Germany.
ORCID-Profile: Thomas Franz Krebs, Ulf Lorenzen, Lidya-Olgu Durmaz, Ewan Brownlee, Andreas Meinzer, Robert Bergholz
Förderung
- CSRD VA
- CMR Surgical (CMR Surgical Ltd, 194 Science Park, Milton Road, Cambridge, UK)
Zitieren
APA
Krebs, T. F., Kayser, T., Lorenzen, U., Grünewald, M., Kayser, M., Saltner, A., Durmaz, L. O., Reese, L. J., Brownlee, E., Reischig, K., Baastrup, J., Meinzer, A., Kalz, A., Becker, T., & Bergholz, R. (2023). Evaluation of the Versius Robotic System for Infant Surgery—A Study in Piglets of Less than 10 kg Body Weight. Children, 10(5), 831. https://doi.org/10.3390/children10050831
Vancouver
Krebs TF, Kayser T, Lorenzen U, Grünewald M, Kayser M, Saltner A, et al. Evaluation of the Versius Robotic System for Infant Surgery—A Study in Piglets of Less than 10 kg Body Weight. Children. 2023;10(5):831. doi:10.3390/children10050831
BibTeX
@article{Krebs2023,
author = {Krebs, Thomas Franz and Kayser, Timo and Lorenzen, Ulf and Grünewald, Matthias and Kayser, Marit and Saltner, Anna and Durmaz, Lidya-Olgu and Reese, Lina Johanna and Brownlee, Ewan and Reischig, Katja and Baastrup, Jonas and Meinzer, Andreas and Kalz, Almut and Becker, Thomas and Bergholz, Robert},
title = {Evaluation of the Versius Robotic System for Infant Surgery—A Study in Piglets of Less than 10 kg Body Weight},
journal = {Children},
year = {2023},
volume = {10},
number = {5},
pages = {831},
doi = {10.3390/children10050831}
}
EUVIMED