EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Multi-Frequency Electrocochleography and Electrode Scan to Identify Electrode Insertion Trauma during Cochlear Implantation

Aniket A Saoji, Madison K Graham, Weston J Adkins, Kanthaiah Koka, Matthew L Carlson, Brian A Neff, Colin L W Driscoll, Douglas C Fitzpatrick

Brain Sciences · 15.02.2023 · 13 (2) · 330

DOI: 10.3390/brainsci13020330 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Brain Sciences, online erschienen am 15.02.2023. Verfasst von Aniket A Saoji und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otolaryngology Head and Neck Surgery, Mayo Clinic, Rochester, MN 55905, USA. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Intraoperative electrocochleography (ECOG) is performed using a single low-frequency acoustic stimulus (e.g., 500 Hz) to monitor cochlear microphonics (CM) during cochlear implant (CI) electrode insertion. A decrease in CM amplitude is commonly associated with cochlear trauma and is used to guide electrode placement. However, advancement of the recording electrode beyond the sites of CM generation can also lead to a decrease in CM amplitude and is sometimes interpreted as cochlear trauma, resulting in unnecessary electrode manipulation and increased risk of cochlear trauma during CI electrode placement. In the present study, multi-frequency ECOG was used to monitor CM during CI electrode placement. The intraoperative CM tracings were compared with electrode scan measurements, where CM was measured for each of the intracochlear electrodes. Comparison between the peak CM amplitude measured during electrode placement and electrode scan measurements was used to differentiate between different mechanisms for decrease in CM amplitude during CI electrode insertion. Analysis of the data shows that both multi-frequency electrocochleography and electrode scan could potentially be used to differentiate between different mechanisms for decreasing CM amplitude and providing appropriate feedback to the surgeon during CI electrode placement.

Schlagwörter

Stichwörter der Autor:innen

cochlear implantscochlear microphonicselectrocochleographyelectrode scan

Autor:innen und Institutionen

  • Department of Otolaryngology Head and Neck Surgery, Mayo Clinic, Rochester, MN 55905, USA.
  • Department of Research and Technology, Advanced Bionics, Valencia, CA 91355, USA.
  • Department of Otolaryngology-Head and Neck Surgery, University of North Carolina at Chapel Hill, Chapel Hill, NC 27599, USA.

Zitieren

APA

Saoji, A. A., Graham, M. K., Adkins, W. J., Koka, K., Carlson, M. L., Neff, B. A., Driscoll, C. L. W., & Fitzpatrick, D. C. (2023). Multi-Frequency Electrocochleography and Electrode Scan to Identify Electrode Insertion Trauma during Cochlear Implantation. Brain Sciences, 13(2), 330. https://doi.org/10.3390/brainsci13020330

Vancouver

Saoji AA, Graham MK, Adkins WJ, Koka K, Carlson ML, Neff BA, et al. Multi-Frequency Electrocochleography and Electrode Scan to Identify Electrode Insertion Trauma during Cochlear Implantation. Brain Sciences. 2023;13(2):330. doi:10.3390/brainsci13020330

BibTeX
@article{Saoji2023,
  author = {Saoji, Aniket A and Graham, Madison K and Adkins, Weston J and Koka, Kanthaiah and Carlson, Matthew L and Neff, Brian A and Driscoll, Colin L W and Fitzpatrick, Douglas C},
  title = {Multi-Frequency Electrocochleography and Electrode Scan to Identify Electrode Insertion Trauma during Cochlear Implantation},
  journal = {Brain Sciences},
  year = {2023},
  volume = {13},
  number = {2},
  pages = {330},
  doi = {10.3390/brainsci13020330}
}