EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Evaluation of an impedance-based method to monitor the insertion of the electrode array during cochlear implantation

Timo M Gottfried, Pablo Galeazzi, Aline Föger, Daniel Dejaco, Andrea Tröger, Natalie Fischer, Veronika Innerhofer, Francesco Di Trapani, Nora Weiss, Josef Seebacher, Angelika Dierker, Joachim Schmutzhard

European Archives of Oto-Rhino-Laryngology · 2024-08 · 281 (8) · 4121-4131

DOI: 10.1007/s00405-024-08584-2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Archives of Oto-Rhino-Laryngology, online erschienen am 02.04.2024. Verfasst von Timo M Gottfried und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria. Hauptthemen laut MEDLINE: Cochlear Implantation, Electric Impedance, Feasibility Studies, Cochlear Implants. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Innsbruck and Medical University of Innsbruck.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Cochlear implantation is a prevalent remedy for severe-to-profound hearing loss. Optimising outcomes and hearing preservation, and minimising insertion trauma, require precise electrode placement. Objective monitoring during the insertion process can provide valuable insights and enhance surgical precision. This study assesses the feasibility and performance of an impedance-based method for monitoring electrode insertion, compared to the surgeon’s feedback.

Methods

The study utilised the Insertion Monitoring Tool (IMT) research software, allowing for real-time measurement of impedance and evoked compound action potential (eCAP) during electrode insertion in 20 patient implantations. This enabled an impedance-based method to continuously assess the status of each electrode during the insertion process. The feasibility and performance was evaluated and compared to the surgeon’s feedback approach. eCAP measurements focused merely on feasibility without searching specific responses.

Results

The IMT demonstrated feasibility in measuring real-time impedances and eCAP during the insertion of the electrode array. The impedance-based method exhibited potential for accurately monitoring the insertion depth with a high success rate. However, further development is needed to improve the number of usable contacts.

Conclusions

Objective monitoring with the impedance-based method shows promise as a valuable tool to enhance the precision of cochlear implant electrode insertion respecting insertion distance estimation. The IMT research software proved feasible in recording real-time impedances and eCAP during electrode insertion. While this impedance-based method exhibits high success rates, further improvements are required to optimise the number of usable contacts. This study highlights the potential of objective monitoring techniques to enhance cochlear implantation outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cochlear ImplantationElectric ImpedanceFeasibility StudiesCochlear ImplantsHumansMaleFemaleMiddle AgedAdultAgedChildChild, PreschoolAdolescentElectrodes, ImplantedEvoked Potentials, AuditoryYoung AdultMonitoring, Intraoperative

Stichwörter der Autor:innen

Cochlear implantElectrode impedanceInsertion speedReal-time measurement

Autor:innen und Institutionen

  • Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria.
  • MED-EL Medical Electronics, Worldwide Headquarters, Fürstenweg 77a, Innsbruck, Tyrol, Austria.
  • Department of Otorhinolaryngology, Head and Neck Surgery, Medical University of Bochum, North Rhine-Westphalia, Bleichstraße 15, 44787, Bochum, Germany.
  • Department of Otorhinolaryngology, TUM School of Medicine, Klinikum rechts der Isar, Technical University of Munich, Munich, Germany.
  • Department of Translational Neurosciences, Faculty of Medicine and Health Sciences, University of Antwerp, Antwerp, Belgium.
  • Departement of Hearing, Speech and Voice Disorders, Medical University of Innsbruck, Anichstr. 35, 6020, Innsbruck, Tyrol, Austria.

Förderung

  • University of Innsbruck and Medical University of Innsbruck

Zitieren

APA

Gottfried, T. M., Galeazzi, P., Föger, A., Dejaco, D., Tröger, A., Fischer, N., Innerhofer, V., Di Trapani, F., Weiss, N., Seebacher, J., Dierker, A., & Schmutzhard, J. (2024). Evaluation of an impedance-based method to monitor the insertion of the electrode array during cochlear implantation. European Archives of Oto-Rhino-Laryngology, 281(8), 4121-4131. https://doi.org/10.1007/s00405-024-08584-2

Vancouver

Gottfried TM, Galeazzi P, Föger A, Dejaco D, Tröger A, Fischer N, et al. Evaluation of an impedance-based method to monitor the insertion of the electrode array during cochlear implantation. European Archives of Oto-Rhino-Laryngology. 2024;281(8):4121-4131. doi:10.1007/s00405-024-08584-2

BibTeX
@article{Gottfried2024,
  author = {Gottfried, Timo M and Galeazzi, Pablo and Föger, Aline and Dejaco, Daniel and Tröger, Andrea and Fischer, Natalie and Innerhofer, Veronika and Di Trapani, Francesco and Weiss, Nora and Seebacher, Josef and Dierker, Angelika and Schmutzhard, Joachim},
  title = {Evaluation of an impedance-based method to monitor the insertion of the electrode array during cochlear implantation},
  journal = {European Archives of Oto-Rhino-Laryngology},
  year = {2024},
  volume = {281},
  number = {8},
  pages = {4121--4131},
  doi = {10.1007/s00405-024-08584-2}
}
Kostenlos starten →