EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Automated Intracellular Immunofluorescence Staining Enabled by Magnetic 3D Mixing in a Modular Microfluidic Platform

Zhengyi Zhang, Mengyu Wang, Runtao Zhong, Yingbo Zhao, Yeqing Sun iD

Biosensors · 2026

DOI: 10.3390/bios16020120 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2026. Verfasst von Zhengyi Zhang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Environmental Systems Biology, College of Environmental Science and Engineering, Dalian Maritime University, Dalian 116026, China. Hauptthemen laut MEDLINE: Microfluidics, Fluorescent Antibody Technique, Biosensing Techniques. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the Space Medical Experiment Project of China Manned Space Program, the Space Application System of China Manned Space Program, the Fundamental Research Funds for the Central Universities.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Traditional sample preparation for flow cytometry is often labor-intensive, operator-dependent, and reagent-consuming, limiting its suitability for automated and point-of-care biosensing applications. To address these challenges, this study presents a functional modular microfluidic system integrating immunomagnetic beads (IMBs) to enable automated intracellular immunofluorescence (IF) staining. The modular microfluidic platform is enabled by a dynamically actuated three-dimensional magnetic field that couples with IMBs within a microfluidic reaction chamber, requiring only one-dimensional magnet translation to induce effective three-dimensional bead motion. This magnetic–chip cooperative strategy significantly enhances microscale mixing and cell capture, facilitating automated immunostaining of the radiation biomarker in CD4+ cells. Finite element simulations were employed to guide magnetic field design by analyzing magnetic force distributions and identifying key parameters, including magnet material, size, spatial arrangement, and chip–magnet distance. Experimental validation using CD4+ cell capture confirmed the effectiveness of the magnetic mixing strategy, revealing ∇B·B as the critical design parameter. An N52 NdFeB magnet (6 mm diameter, 10 mm height) positioned within 2.2 mm of the chamber centerline stably retained IMBs at flow rates below 200 µL/min. Under optimized conditions (magnet translation speed of 8 mm/s and a 15 min mixing duration), a maximum cell capture efficiency of 86% was achieved. Subsequent automated γH2AX IF staining demonstrated a strong linear dose–response relationship (R2 > 0.9) in mean fluorescence intensity. This study demonstrates a robust and scalable strategy for automating complex IF staining workflows, highlighting the potential of magnetic-field-assisted microfluidic platforms for biosensing applications requiring reliable intracellular biomarker detection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MicrofluidicsFluorescent Antibody TechniqueBiosensing TechniquesHumansMicrofluidic Analytical TechniquesMagnetic FieldsStaining and LabelingFlow Cytometry

Stichwörter der Autor:innen

immunofluorescence (IF) stainingimmunomagnetic beads (IMBs)magnetic field simulationmicrofluidic

Autor:innen und Institutionen

  • Institute of Environmental Systems Biology, College of Environmental Science and Engineering, Dalian Maritime University, Dalian 116026, China.

ORCID-Profile: Yeqing Sun

Förderung

  • the Space Medical Experiment Project of China Manned Space Program
  • the Space Application System of China Manned Space Program
  • the Fundamental Research Funds for the Central Universities
  • Space Medical Experiment Project of China Manned Space Program
  • Space Application System of China Manned Space Program
  • Fundamental Research Funds for the Central Universities

Zitieren

APA

Zhang, Z., Wang, M., Zhong, R., Zhao, Y., & Sun, Y. (2026). Automated Intracellular Immunofluorescence Staining Enabled by Magnetic 3D Mixing in a Modular Microfluidic Platform. Biosensors. https://doi.org/10.3390/bios16020120

Vancouver

Zhang Z, Wang M, Zhong R, Zhao Y, Sun Y. Automated Intracellular Immunofluorescence Staining Enabled by Magnetic 3D Mixing in a Modular Microfluidic Platform. Biosensors. 2026. doi:10.3390/bios16020120

BibTeX
@article{Zhang2026,
  author = {Zhang, Zhengyi and Wang, Mengyu and Zhong, Runtao and Zhao, Yingbo and Sun, Yeqing},
  title = {Automated Intracellular Immunofluorescence Staining Enabled by Magnetic 3D Mixing in a Modular Microfluidic Platform},
  journal = {Biosensors},
  year = {2026},
  doi = {10.3390/bios16020120}
}
Kostenlos starten →