EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study

Tahsin Nairuz iD, Young-Suk Hwang iD, Min-Yong Kwon iD, Jae Hyun Kim, Sae Min Kwon, Hyuck-Jun Yoon iD, Seung-Ho Hur, Joonho Chung iD, Woo-Jin Kim, Sang-Hyun An, Jun Sik Kim, Jong-Ha Lee iD und 1 weitere

Biosensors · 2025

DOI: 10.3390/bios15110755 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Tahsin Nairuz und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Keimyung University, Daegu 42601, Republic of Korea. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Catheters. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the Korea Health Technology R&D Project through the Korea Health Industry Development In-stitute (KHIDI), funded by the Ministry of Health & Welfare, Republic…, Korea Basic Science Institute, Ministry of Education.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The application of intravascular optical coherence tomography (IV-OCT) in neurovascular interventions is constrained by the mechanical inadequacy of conventional catheters in navigating the complex intracranial vasculature. To address this, we modified a coronary IV-OCT catheter, enhancing its mechanical performance for neurovascular applications. The modified catheter featured a 300 mm over-the-microwire segment and a dual-structured shaft (distal 50 mm nonbraided, proximal 250 mm braided) to improve trackability and pushability. We compared the modified and conventional catheters using a benchtop model with a simulated vessel path and an in vivo swine model. Trackability and pushability were quantitatively measured using resistance (N) and advancement distance (mm) in the simulated path. In the animal model, indirect performance metrics included the catheter tension angle (CTA) and pass of catheter (POC) through the fourth curvature of the external carotid artery (ECA). The modified catheter demonstrated superior pushability (172.9 ± 1.96 mm vs. 127.9 ± 2.86 mm, p < 0.05) and increased resistance (1.47 ± 0.036 N vs. 0.69 ± 0.032 N, p < 0.05). In vivo analysis further showed a significantly greater CTA (115.8 ± 8.5° vs. 77.6 ± 10.3°, p < 0.05) and higher POC success rate (83.3% vs. 11.1%, p < 0.05). These results indicate that the modified coronary IV-OCT catheter offers enhanced mechanical performance, suggesting its potential for safe and effective use in neurovascular procedures.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceCathetersAnimalsSwineEquipment Design

Stichwörter der Autor:innen

catheter modificationintravascular optical coherence tomographymechanical performanceneurovascular interventionstrackability and pushability

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Keimyung University, Daegu 42601, Republic of Korea.
  • Department of Neurosurgery, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.
  • Department of Cardiology, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.
  • Department of Neurosurgery, Yonsei University College of Medicine, Seoul 6273, Republic of Korea.
  • Biomedical Manufacturing Technology Center (BMTC), Korea Institute of Industrial Technology (KITECH), Yeongcheon 38822, Republic of Korea.
  • Preclinical Research Center, Daegu-Gyeongbuk Medical Innovation Foundation (KMedi Hub), Daegu 41061, Republic of Korea.

ORCID-Profile: Tahsin Nairuz, Young-Suk Hwang, Min-Yong Kwon, Hyuck-Jun Yoon, Joonho Chung, Jong-Ha Lee, Chang-Hyun Kim

Förderung

  • the Korea Health Technology R&D Project through the Korea Health Industry Development In-stitute (KHIDI), funded by the Ministry of Health & Welfare, Republic…
  • Korea Basic Science Institute
  • Ministry of Education
  • Regional Innovation System & Education (RISE) program
  • Daegu RISE Center
  • Daegu, Republic of Korea
  • Korea Medical Device Development Fund
  • Korea government

Zitieren

APA

Nairuz, T., Hwang, Y. S., Kwon, M. Y., Kim, J. H., Kwon, S. M., Yoon, H. J., Hur, S. H., Chung, J., Kim, W. J., An, S. H., Kim, J. S., Lee, J. H., & Kim, C. H. (2025). Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study. Biosensors. https://doi.org/10.3390/bios15110755

Vancouver

Nairuz T, Hwang YS, Kwon MY, Kim JH, Kwon SM, Yoon HJ, et al. Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study. Biosensors. 2025. doi:10.3390/bios15110755

BibTeX
@article{Nairuz2025,
  author = {Nairuz, Tahsin and Hwang, Young-Suk and Kwon, Min-Yong and Kim, Jae Hyun and Kwon, Sae Min and Yoon, Hyuck-Jun and Hur, Seung-Ho and Chung, Joonho and Kim, Woo-Jin and An, Sang-Hyun and Kim, Jun Sik and Lee, Jong-Ha and Kim, Chang-Hyun},
  title = {Design Optimization and Mechanical Performance Evaluation of a Modified Coronary IV-OCT Catheter Adapted for Intracranial Navigation: A Preclinical Study},
  journal = {Biosensors},
  year = {2025},
  doi = {10.3390/bios15110755}
}
Kostenlos starten →