Strain Plethysmography Using a Hermetically Sealed MEMS Strain Sensor

Xinyu Jiang, Brian Sang iD, Haoran Wen, Gregory Junek, Jin-Woo Park, Farrokh Ayazi

Biosensors · 2025

DOI: 10.3390/bios15050325 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Xinyu Jiang und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Electrical and Computer Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USA. Hauptthemen laut MEDLINE: Plethysmography, Micro-Electrical-Mechanical Systems, Biosensing Techniques. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

We present a hermetically sealed capacitive microelectromechanical system (MEMS) strain sensor designed for arterial pulse waveform extraction using the strain plethysmography (SPG) modality. The MEMS strain sensor features a small form factor of 3.3 mm × 3.3 mm × 1 mm, leverages a nano-gap fabrication process to improve the sensitivity, and uses a differential sensing mechanism to improve the linearity and remove the common mode drift. The MEMS strain sensor is interfaced with an application-specific integrated circuit (ASIC) to form a compact strain sensing system. This system exhibits a high strain sensitivity of 316 aF/µε, a gauge factor (GF) of 35, and a strain sensing resolution of 1.26 µε, while maintaining a linear range exceeding 700 µε. SPG signals have been reliably captured at both the fingertip and wrist using the MEMS strain sensor with high signal quality, preserving various photoplethysmography (PPG) features. Experimental results demonstrate that heart rate (HR) and heart rate variability (HRV) can be estimated from the SPG signal collected at the fingertip and wrist using the sensor with an accuracy of over 99%. Pulse arrival time (PAT) and pulse transit time (PTT) have been successfully extracted using the sensor together with a MEMS seismometer, showcasing its potential for ambulatory BP monitoring (ABPM) application.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PlethysmographyMicro-Electrical-Mechanical SystemsBiosensing TechniquesHumansHeart RatePulse Wave Analysis

Stichwörter der Autor:innen

Blood PressureMEMSarterial pulse waveformcardiovascular diseasepulse transit timestrain sensors

Autor:innen und Institutionen

  • School of Electrical and Computer Engineering, Georgia Institute of Technology, Atlanta, GA 30332, USA.
  • StethX Microsystems Inc., Atlanta, GA 30308, USA.

ORCID-Profile: Brian Sang

Zitieren

APA

Jiang, X., Sang, B., Wen, H., Junek, G., Park, J. W., & Ayazi, F. (2025). Strain Plethysmography Using a Hermetically Sealed MEMS Strain Sensor. Biosensors. https://doi.org/10.3390/bios15050325

Vancouver

Jiang X, Sang B, Wen H, Junek G, Park JW, Ayazi F. Strain Plethysmography Using a Hermetically Sealed MEMS Strain Sensor. Biosensors. 2025. doi:10.3390/bios15050325

BibTeX
@article{Jiang2025,
  author = {Jiang, Xinyu and Sang, Brian and Wen, Haoran and Junek, Gregory and Park, Jin-Woo and Ayazi, Farrokh},
  title = {Strain Plethysmography Using a Hermetically Sealed MEMS Strain Sensor},
  journal = {Biosensors},
  year = {2025},
  doi = {10.3390/bios15050325}
}
Kostenlos starten →