A Novel Adaptive Independent Component Analysis Method for Multi-Channel Optically Pumped Magnetometers’ Magnetocardiography Signals
Biosensors · 2025
DOI: 10.3390/bios15040243 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, 2025. Verfasst von Shuang Liang und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Instrumentation Science and Optoelectronics Engineering, Beihang University, Beijing 100191, China. Hauptthemen laut MEDLINE: Magnetocardiography, Magnetometry. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key Research and Development Program of China.
Abstract (Originalsprache)
With the gradual maturation of optically pumped magnetometer (OPM) technology, the use of OPMs to acquire weak magnetocardiography (MCG) signals has started to gain widespread application. Due to the complexity of magnetic environments, MCG signals are often subject to interference from various unknown sources. Independent component analysis (ICA) is one of the most widely used methods for blind source separation. However, in practical applications, the numbers of retained components and filtering components are often selected manually, relying on subjective experience. This study proposes an adaptive ICA method that estimates the signal-to-noise ratio (SNR) before processing to determine the number of components and selects heartbeat-related components based on their characteristic indicators. The method was validated using phantom experiments and MCG data in a 128-channel OPM-MCG system. In the human subject experiment, the array output SNR reached 31.8 dB, and the processing time was significantly reduced to 1/38 of the original. The proposed method outperformed traditional techniques in terms of its ability to identify artifacts and efficiency in this regard, providing strong support for the broader clinical application of OPM-MCG.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Instrumentation Science and Optoelectronics Engineering, Beihang University, Beijing 100191, China.
- School of Electronics, Peking University, Beijing 100871, China.
- PKU-Nanjing Institute of Translational Medicine, Nanjing Raygen Health, Nanjing 210031, China.
- Beijing Laboratory of Biomedical Imaging, Beijing Municipal Education Commission, Beijing 100085, China.
- National Biomedical Imaging Center, State Key Laboratory of Membrane Biology, Peking-Tsinghua Center for Life Sciences, College of Future Technology, Peking University, Beijing 100871, China.
ORCID-Profile: Shuang Liang, Jiahe Qi, Junhuai He, Yikang Jia, Aimin Wang, Ting Zhao, Chaoliang Wei, Hongchen Jiao, Lishuang Feng, Heping Cheng
Förderung
- National Key Research and Development Program of China
Zitieren
APA
Liang, S., Qi, J., He, J., Jia, Y., Wang, A., Zhao, T., Wei, C., Jiao, H., Feng, L., & Cheng, H. (2025). A Novel Adaptive Independent Component Analysis Method for Multi-Channel Optically Pumped Magnetometers’ Magnetocardiography Signals. Biosensors. https://doi.org/10.3390/bios15040243
Vancouver
Liang S, Qi J, He J, Jia Y, Wang A, Zhao T, et al. A Novel Adaptive Independent Component Analysis Method for Multi-Channel Optically Pumped Magnetometers’ Magnetocardiography Signals. Biosensors. 2025. doi:10.3390/bios15040243
BibTeX
@article{Liang2025,
author = {Liang, Shuang and Qi, Jiahe and He, Junhuai and Jia, Yikang and Wang, Aimin and Zhao, Ting and Wei, Chaoliang and Jiao, Hongchen and Feng, Lishuang and Cheng, Heping},
title = {A Novel Adaptive Independent Component Analysis Method for Multi-Channel Optically Pumped Magnetometers’ Magnetocardiography Signals},
journal = {Biosensors},
year = {2025},
doi = {10.3390/bios15040243}
}