EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Development of a Brain Catheter for Optical Coherence Tomography in Advanced Cerebrovascular Diagnostics

Tae-Mi Jung iD, Tahsin Nairuz iD, Chang-Hyun Kim, Jong-Ha Lee iD

Biosensors · 06.03.2025 · 15 (3) · 170

DOI: 10.3390/bios15030170 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors, online erschienen am 06.03.2025. Verfasst von Tae-Mi Jung und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Keimyung University, Daegu 42601, Republic of Korea. Hauptthemen laut MEDLINE: Tomography, Optical Coherence, Catheters, Brain, Cerebrovascular Disorders. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Trade, Industry and Energy through the Research and Development Pro-gram, Korea government (MSIT and Daegu Metropolitan City), National Research Foundation of Korea.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Optical coherence tomography (OCT) has been extensively utilized in cardiovascular diagnostics due to its high resolution, rapid imaging capabilities; however, its adaptation for cerebrovascular applications remains constrained by the narrow, tortuous anatomical structure of cerebral vessels. To address these limitations, this study introduces a cerebrovascular-specific OCT (bOCT) catheter, an advanced adaptation of the cardiovascular OCT (cOCT) catheter, with significant structural modifications for improved access to brain blood vessels. The bOCT catheter incorporates a braided wire within a braided tube, strategically reinforcing axial strength. The distal shaft was reconfigured as a single-lumen structure, facilitating unified movement of the rotating fiber optic core and guidewire, thereby reducing guidewire bending and augmenting force transmission stability. Additionally, the anterior protrusion was removed and replaced with a dual-lumen configuration, significantly enhancing lesion accessibility. The bOCT catheter’s performance was validated in a 3D physical model and an animal model, demonstrating pronounced enhancements in flexibility, pushability, and navigability. Notably, the pushability through curved flow paths significantly improved, enhancing access to cerebral blood vessels. Therefore, this innovation promises to revolutionize cerebrovascular diagnostics with high-resolution imaging suited to the complex brain vasculature, setting a new standard in intravascular imaging technology.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, Optical CoherenceCathetersBrainCerebrovascular DisordersAnimalsHumans

Stichwörter der Autor:innen

cardiovascular OCT cathetercerebrovascular OCT cathetercerebrovascular diagnosticsoptical coherence tomography

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Keimyung University, Daegu 42601, Republic of Korea.
  • Department of Neurosurgery, Keimyung University Dongsan Hospital, Daegu 42601, Republic of Korea.

ORCID-Profile: Tae-Mi Jung, Tahsin Nairuz, Jong-Ha Lee

Förderung

  • Ministry of Trade, Industry and Energy through the Research and Development Pro-gram
  • Korea government (MSIT and Daegu Metropolitan City)
  • National Research Foundation of Korea
  • Ministry of Health & Welfare, Republic of Korea
  • Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health & Welfare, the Ministry of Food and D…
  • Ministry of Education

Zitieren

APA

Jung, T. M., Nairuz, T., Kim, C. H., & Lee, J. H. (2025). Development of a Brain Catheter for Optical Coherence Tomography in Advanced Cerebrovascular Diagnostics. Biosensors, 15(3), 170. https://doi.org/10.3390/bios15030170

Vancouver

Jung TM, Nairuz T, Kim CH, Lee JH. Development of a Brain Catheter for Optical Coherence Tomography in Advanced Cerebrovascular Diagnostics. Biosensors. 2025;15(3):170. doi:10.3390/bios15030170

BibTeX
@article{Jung2025,
  author = {Jung, Tae-Mi and Nairuz, Tahsin and Kim, Chang-Hyun and Lee, Jong-Ha},
  title = {Development of a Brain Catheter for Optical Coherence Tomography in Advanced Cerebrovascular Diagnostics},
  journal = {Biosensors},
  year = {2025},
  volume = {15},
  number = {3},
  pages = {170},
  doi = {10.3390/bios15030170}
}
Kostenlos starten →