Enhanced Sensitivity Mach–Zehnder Interferometer-Based Tapered-in-Tapered Fiber-Optic Biosensor for the Immunoassay of C-Reactive Protein
Biosensors · 06.02.2025 · 15 (2) · 90
DOI: 10.3390/bios15020090 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, online erschienen am 06.02.2025. Verfasst von Lei Xiao und 8 weiteren Autor:innen. Beteiligte Einrichtung: College of Physics and Optoelectronic Engineering, Shenzhen University, Shenzhen 518060, China. Hauptthemen laut MEDLINE: Biosensing Techniques, C-Reactive Protein, Interferometry, Fiber Optic Technology. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, High-end Talent Scientific Research Startup Project, Natural Science Foundation of Guangdong.
Abstract (Originalsprache)
A Mach–Zehnder interferometer-based tapered-in-tapered fiber-optic biosensor was introduced in this paper. By integrating a micro-tapered fiber into a single tapered fiber structure, the design enhances sensitivity, signal-to-noise ratio, and resolution capability, while reducing the length of the sensing fiber. Through simulation analysis, it was found that the tapered-in-tapered fiber significantly improved the refractive index detection sensitivity by exciting a stronger evanescent field effect. The experimental comparison between the tapered-in-tapered fiber and traditional tapered fiber showed a 1.7-fold increase in sensitivity, reaching 3266.78 nm/RIU within the refractive index range of 1.3326 to 1.3414. Furthermore, to expand its application prospects in the biomedical field, glutaraldehyde cross-linking technology was used to immobilize C-reactive protein (CRP) antibodies on the surface of the tapered-in-tapered fiber, successfully creating a biosensing platform for the specific recognition of CRP. The experimental results demonstrate that this novel biosensor can rapidly and accurately detect CRP molecules at different concentrations with a detection limit of 0.278 μg/mL, and that it exhibits good selectivity and repeatability. This tapered-in-tapered fiber-optic biosensor provides new insights into the development of high-performance fiber-optic immunosensors and shows broad application potential in immunology research and early disease diagnosis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- College of Physics and Optoelectronic Engineering, Shenzhen University, Shenzhen 518060, China.
- Shenzhen Engineering Laboratory for Optical Fiber Sensors and Networks, Shenzhen 518060, China.
- Shenzhen Tian'an Zhiyuan Sensor Technology Co., Ltd., Shenzhen 518060, China.
- Shenzhen Key Laboratory of Sensor Technology, Shenzhen 518060, China.
- School of Science, The Chinese University of Hong Kong, Shenzhen 518172, China.
ORCID-Profile: Xinghong Chen, Yuzhi Chen
Förderung
- National Natural Science Foundation of China
- High-end Talent Scientific Research Startup Project
- Natural Science Foundation of Guangdong
- Science and Technology plan project of Shenzhen
- LingChuang Research Project of China National Nuclear Corporation
Zitieren
APA
Xiao, L., Chen, X., Li, X., Zhang, J., Wang, Y., Li, D., Hong, X., Shao, Y., & Chen, Y. (2025). Enhanced Sensitivity Mach–Zehnder Interferometer-Based Tapered-in-Tapered Fiber-Optic Biosensor for the Immunoassay of C-Reactive Protein. Biosensors, 15(2), 90. https://doi.org/10.3390/bios15020090
Vancouver
Xiao L, Chen X, Li X, Zhang J, Wang Y, Li D, et al. Enhanced Sensitivity Mach–Zehnder Interferometer-Based Tapered-in-Tapered Fiber-Optic Biosensor for the Immunoassay of C-Reactive Protein. Biosensors. 2025;15(2):90. doi:10.3390/bios15020090
BibTeX
@article{Xiao2025,
author = {Xiao, Lei and Chen, Xinghong and Li, Xuejin and Zhang, Jinghan and Wang, Yan and Li, Dongqing and Hong, Xueming and Shao, Yonghong and Chen, Yuzhi},
title = {Enhanced Sensitivity Mach–Zehnder Interferometer-Based Tapered-in-Tapered Fiber-Optic Biosensor for the Immunoassay of C-Reactive Protein},
journal = {Biosensors},
year = {2025},
volume = {15},
number = {2},
pages = {90},
doi = {10.3390/bios15020090}
}
EUVIMED