Using Temporally and Spatially Resolved Measurements to Improve the Sensitivity of Fluorescence-Based Immunoassays
Biosensors · 28.04.2024 · 14 (5) · 220
DOI: 10.3390/bios14050220 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors, online erschienen am 28.04.2024. Verfasst von Ran Kremer und 4 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Engineering, The Institute of Nanotechnology and Advanced Materials, Bar-Ilan University, Max and Anna Webb Street, Ramat Gan 5290002, Isr…. Hauptthemen laut MEDLINE: Biosensing Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Israel Science Foundation, Ministry of Innovation, Science and Technology, Israel Ministry of Science.
Abstract (Originalsprache)
Detecting low concentrations of biomarkers is essential in clinical laboratories. To improve analytical sensitivity, especially in identifying fluorescently labeled molecules, typical optical detection systems, consisting of a photodetector or camera, utilize time-resolved measurements. Taking a different approach, magnetic modulation biosensing (MMB) is a novel technology that combines fluorescently labeled probes and magnetic particles to create a sandwich assay with the target molecules. By concentrating the target molecules and then using time-resolved measurements, MMB provides the rapid and highly sensitive detection of various biomarkers. Here, we propose a novel signal-processing algorithm that enhances the detection and estimation of target molecules at low concentrations. By incorporating both temporally and spatially resolved measurements using human interleukin-8 as a target molecule, we show that the new algorithm provides a 2–4-fold improvement in the limit of detection and an ~25% gain in quantitative resolution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Faculty of Engineering, The Institute of Nanotechnology and Advanced Materials, Bar-Ilan University, Max and Anna Webb Street, Ramat Gan 5290002, Israel.
ORCID-Profile: Shira Roth, Amos Danielli, Yair Noam
Förderung
- Israel Science Foundation
- Ministry of Innovation, Science and Technology
- Israel Ministry of Science
- Bar Ilan University’s Dangoor Centre
Zitieren
APA
Kremer, R., Roth, S., Bross, A., Danielli, A., & Noam, Y. (2024). Using Temporally and Spatially Resolved Measurements to Improve the Sensitivity of Fluorescence-Based Immunoassays. Biosensors, 14(5), 220. https://doi.org/10.3390/bios14050220
Vancouver
Kremer R, Roth S, Bross A, Danielli A, Noam Y. Using Temporally and Spatially Resolved Measurements to Improve the Sensitivity of Fluorescence-Based Immunoassays. Biosensors. 2024;14(5):220. doi:10.3390/bios14050220
BibTeX
@article{Kremer2024,
author = {Kremer, Ran and Roth, Shira and Bross, Avital and Danielli, Amos and Noam, Yair},
title = {Using Temporally and Spatially Resolved Measurements to Improve the Sensitivity of Fluorescence-Based Immunoassays},
journal = {Biosensors},
year = {2024},
volume = {14},
number = {5},
pages = {220},
doi = {10.3390/bios14050220}
}
EUVIMED