Fachartikel in der Zeitschrift Biosensors, 2023. Verfasst von Yeeun Kim und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Semiconductor & Display Technology, Hallym University, Chuncheon 24252, Republic of Korea. Hauptthemen laut MEDLINE: Biosensing Techniques. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Hallym University Research Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In this study, an impedance biosensor capable of real-time monitoring of the growth and drug reactions using NIH/3T3 cells was fabricated through a semiconductor process. With the fabricated impedance biosensor, the cell growth and drug reaction states are monitored in real-time, showing the validness of the developed biosensor. By using the developed impedance biosensor, we have investigated the capacitance contribution of NIH/3T3 cells existing on electrodes and between electrodes. To compare the capacitance value contributions of the cells on and between electrodes, wide- and narrow-gap electrode patterns are manufactured with 3.7 and 0.3 mm electrode gap spacings, respectively. From the detailed analysis, the capacitance contributions of NIH/3T3 cells existing on electrodes are estimated around less than 20 percent compared to the cells existing between electrodes. In other words, a minimized electrode area with maximized electrode spacing is the promising impedance biosensor design guide for accurate cell capacitance measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesAnimalsMiceElectric ImpedanceElectrodesElectric Capacitance3T3 Cells

Stichwörter der Autor:innen

NIH 3T3 cellcapacitanceimpedance biosensor

Autor:innen und Institutionen

  • School of Semiconductor & Display Technology, Hallym University, Chuncheon 24252, Republic of Korea.
  • Department of Life Science, Hallym University, Chuncheon 24252, Republic of Korea.
  • Center of Nano Convergence Technology, Hallym University, Chuncheon 24252, Republic of Korea.

Förderung

  • Hallym University Research Fund

Zitieren

APA

Kim, Y., Kang, D., Kim, S., Hong, E., & Jang, M. (2023). Capacitance Contribution of NIH/3T3 Cells Existing on and between Electrodes of an Impedance Biosensor. Biosensors. https://doi.org/10.3390/bios13110970

Vancouver

Kim Y, Kang D, Kim S, Hong E, Jang M. Capacitance Contribution of NIH/3T3 Cells Existing on and between Electrodes of an Impedance Biosensor. Biosensors. 2023. doi:10.3390/bios13110970

BibTeX
@article{Kim2023,
  author = {Kim, Yeeun and Kang, Dahyun and Kim, Seokgyu and Hong, Eunchae and Jang, Moongyu},
  title = {Capacitance Contribution of NIH/3T3 Cells Existing on and between Electrodes of an Impedance Biosensor},
  journal = {Biosensors},
  year = {2023},
  doi = {10.3390/bios13110970}
}
Kostenlos starten →