Accelerated Renal T2-Weighted Turbo Spin-Echo Imaging with Deep Learning Reconstruction: Reduced Acquisition Time and Improved Image Quality in Patients with Renal Tumors
Bioengineering · 23.11.2026 · 13 (12) · 1414
DOI: 10.3390/bioengineering13121414 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 23.11.2026. Verfasst von Mengmeng Gao und 6 weiteren Autor:innen. Die Arbeit nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation.
Abstract (Originalsprache)
As MRI demand increases, reducing acquisition time while maintaining image quality is essential for improving examination efficiency. This study aimed to evaluate the clinical feasibility of an accelerated renal T2-weighted turbo spin-echo (T2DL) protocol with deep learning reconstruction, using standard T2-weighted TSE (T2S) as a reference. This retrospective study included 68 consecutive patients with clinically suspected renal tumors who underwent 3-T MRI with both T2S and T2DL. Quantitative image quality was assessed using apparent signal-to-noise ratio (aSNR) and contrast-to-noise ratio (aCNR). Two readers independently assessed six qualitative image quality criteria using a 5-point Likert scale. T2DL reduced acquisition time by 31.6% (2:25 vs. 3:32 min) and yielded higher aSNRs in renal parenchyma, lesions, psoas muscle, and subcutaneous adipose tissue (SAT), along with higher aCNRs in renal parenchyma, lesions, and SAT (all p < 0.001). T2DL also received significantly higher scores for image sharpness, lesion conspicuity, capsule delineation, diagnostic confidence, and overall image quality, with fewer artifacts (all p < 0.05). VGC analysis further favored T2DL over T2S across all six criteria for both readers (AUCVGC, 0.630–0.795; all p < 0.05). These findings support the clinical feasibility of an accelerated T2DL protocol, which reduced acquisition time while improving image quality.
Förderung
- National Natural Science Foundation
Zitieren
APA
Gao, M., Li, S., He, K., Liao, Z., Chen, W., Lee, J. M., & Li, Z. (2026). Accelerated Renal T2-Weighted Turbo Spin-Echo Imaging with Deep Learning Reconstruction: Reduced Acquisition Time and Improved Image Quality in Patients with Renal Tumors. Bioengineering, 13(12), 1414. https://doi.org/10.3390/bioengineering13121414
Vancouver
Gao M, Li S, He K, Liao Z, Chen W, Lee JM, et al. Accelerated Renal T2-Weighted Turbo Spin-Echo Imaging with Deep Learning Reconstruction: Reduced Acquisition Time and Improved Image Quality in Patients with Renal Tumors. Bioengineering. 2026;13(12):1414. doi:10.3390/bioengineering13121414
BibTeX
@article{Gao2026,
author = {Gao, Mengmeng and Li, Shichao and He, Kangwen and Liao, Zhouyan and Chen, Wei and Lee, Jeong Min and Li, Zhen},
title = {Accelerated Renal T2-Weighted Turbo Spin-Echo Imaging with Deep Learning Reconstruction: Reduced Acquisition Time and Improved Image Quality in Patients with Renal Tumors},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {12},
pages = {1414},
doi = {10.3390/bioengineering13121414}
}