Assessment of Reflection-Mode Imaging Performance for Ultrasound Computed Tomography Using Tissue-Mimicking Phantoms

Zhaohui Liu, Di Zhang, Ming Yuchi, Chenggang Zhu, Wu Qiu, Mingyue Ding, Fengqi Niu

Bioengineering · 07.11.2026 · 13 (11) · 1321

DOI: 10.3390/bioengineering13111321 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 07.11.2026. Verfasst von Zhaohui Liu und 6 weiteren Autor:innen. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the National Key Research and Development Program of China, the National Natural Science Foundation of China, the Science and Technology Department of Hubei Province.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Ultrasound computed tomography (USCT) has emerged as a promising multi-modal imaging technique, yet standardized phantoms for validating its reflection-mode performance remain lacking, hindering system calibration and cross-platform comparison. In this study, we designed and fabricated a suite of agar-based tissue-mimicking phantoms dedicated to quantitative quality assessment of ring-array USCT reflection imaging. The phantom set consists of three individual modules for evaluating spatial resolution, slice thickness and three-dimensional volumetric accuracy. The acoustic properties of the tissue-mimicking material were characterized for speed of sound and attenuation and complied with standard ultrasound specifications. All experiments were implemented on an NMPA-approved 2048-element ring-array USCT system using synthetic transmit aperture imaging. Experimental results revealed obvious resolution anisotropy of the ring array, with lateral resolution outperforming axial resolution. The measured slice thickness decreased gradually as the radial position from the ring center decreased, and volumetric measurement errors of embedded ovoid targets were controlled within 3.65%. This phantom suite provides quantitative assessment of the principal reflection-mode imaging metrics on one ring-array USCT system; repeatability under phantom repositioning was not evaluated. These results demonstrate the feasibility of the proposed phantoms on a single ring-array system; validation across other USCT architectures remains to be performed.

Förderung

  • the National Key Research and Development Program of China
  • the National Natural Science Foundation of China
  • the Science and Technology Department of Hubei Province

Zitieren

APA

Liu, Z., Zhang, D., Yuchi, M., Zhu, C., Qiu, W., Ding, M., & Niu, F. (2026). Assessment of Reflection-Mode Imaging Performance for Ultrasound Computed Tomography Using Tissue-Mimicking Phantoms. Bioengineering, 13(11), 1321. https://doi.org/10.3390/bioengineering13111321

Vancouver

Liu Z, Zhang D, Yuchi M, Zhu C, Qiu W, Ding M, et al. Assessment of Reflection-Mode Imaging Performance for Ultrasound Computed Tomography Using Tissue-Mimicking Phantoms. Bioengineering. 2026;13(11):1321. doi:10.3390/bioengineering13111321

BibTeX
@article{Liu2026,
  author = {Liu, Zhaohui and Zhang, Di and Yuchi, Ming and Zhu, Chenggang and Qiu, Wu and Ding, Mingyue and Niu, Fengqi},
  title = {Assessment of Reflection-Mode Imaging Performance for Ultrasound Computed Tomography Using Tissue-Mimicking Phantoms},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {11},
  pages = {1321},
  doi = {10.3390/bioengineering13111321}
}
Kostenlos starten →