A Constraint-Driven Automated Framework for Optimizing Multi-Tool Fiducial Configurations in Surgical Navigation
Bioengineering · 2026-07-08 · 13 (7) · 786
DOI: 10.3390/bioengineering13070786 ↗At a glance
Journal article in Bioengineering, published online on 2026-07-08. Written by Yuhui Wang and 4 co-authors. Institution: Key Laboratory of Mechanism Theory and Equipment Design of Ministry of Education, Tianjin University, Tianjin 300350, China. 28 references. The full text is freely available (open access, CC BY 4.0). Funded by: Tianjin Major Science and Technology Project on Artificial Intelligence, Guizhou Provincial Science and Technology Support Project, Seed Foundation of Tianjin University.
Abstract (original language)
The accuracy of optical tracking tools is crucial for surgical navigation. While commercial tools are reliable, their proprietary design knowledge limits accessibility and adaptability for specialized clinical and research applications. This study introduces an open, reproducible optimization framework based on point-based rigid registration theory, defining a unified pose estimation deviation metric and deriving its analytical expression for both expectation and variance. The approach incorporates constraints for intra-group uniqueness and inter-group compatibility, using exhaustive configuration generation and geometric evaluation to rank designs by predicted accuracy. Numerical simulations confirmed the derived formula, with under 5% average prediction error for the expectation and strong agreement for the variance. Optimized four-fiducial tools were compared to commercial references via tip calibration, distance measurement, and registration tests. Most optimized tools (75%) achieved accuracy comparable to or modestly better than commercial tools (e.g., tip calibration 0.22 mm vs. 0.28 mm, distance measurement 0.18 mm vs. 0.20 mm, FRE 0.13 mm vs. 0.16 mm, TRE 0.48 mm vs. 0.51 mm). All four experiments showed a strong positive correlation between the theoretical metric and measured error (Pearson’s r>0.98, all p
Subjects
Author keywords
Authors and institutions
- Key Laboratory of Mechanism Theory and Equipment Design of Ministry of Education, Tianjin University, Tianjin 300350, China.
- International Institute for Innovative Design and Intelligent Manufacturing, Tianjin University, Shaoxing 312000, China.
ORCID profiles: Chuanba Liu
Funding
- Tianjin Major Science and Technology Project on Artificial Intelligence
- Guizhou Provincial Science and Technology Support Project
- Seed Foundation of Tianjin University
Cite
APA
Wang, Y., Liu, C., Wang, Y., Yang, B., & Sun, T. (2026). A Constraint-Driven Automated Framework for Optimizing Multi-Tool Fiducial Configurations in Surgical Navigation. Bioengineering, 13(7), 786. https://doi.org/10.3390/bioengineering13070786
Vancouver
Wang Y, Liu C, Wang Y, Yang B, Sun T. A Constraint-Driven Automated Framework for Optimizing Multi-Tool Fiducial Configurations in Surgical Navigation. Bioengineering. 2026;13(7):786. doi:10.3390/bioengineering13070786
BibTeX
@article{Wang2026,
author = {Wang, Yuhui and Liu, Chuanba and Wang, Yifei and Yang, Boyu and Sun, Tao},
title = {A Constraint-Driven Automated Framework for Optimizing Multi-Tool Fiducial Configurations in Surgical Navigation},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {7},
pages = {786},
doi = {10.3390/bioengineering13070786}
}
EUVIMED