EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

A Study of Coronary Stent Thrombosis Growth

Ivan Fernney Ibanez Aguilar iD, Marcos Beltrão Paraizo, Pedro Soares Teixeira iD, Adriano Camargo Carneiro iD, Bruno Alvares de Azevedo iD, Angela Ourivio Nieckele iD, Carlos Eduardo Rochitte iD

Bioengineering · 2026-07-07 · 13 (7) · 779

DOI: 10.3390/bioengineering13070779 ↗

At a glance

Journal article in Bioengineering, published online on 2026-07-07. Written by Ivan Fernney Ibanez Aguilar and 6 co-authors. Institution: Department of Mechanical Engineering, Pontifical Catholic University of Rio de Janeiro, Rio de Janeiro 22451-900, RJ, Brazil. 37 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

In this study, the impact of thrombus growth at the left coronary ostium—resulting from a misplaced stent—was investigated through numerical simulations. Blood flow from the aortic root into the coronary arteries was modeled for several potential thrombus sizes to assess changes in pressure and shear stress caused by thrombus development. The results showed that even in the most severe case examined (with the thrombus growing three times its original size), the obstruction to the coronary cross-sectional area was not sufficient to produce a significant pressure drop or wall shear stress increases. This outcome is primarily attributed to the thrombus’s location at the tip of the stent; even with considerable elongation, it did not significantly narrow the coronary artery’s lumen. While the specific configurations studied did not result in critical obstruction, it is important to note that other thrombus shapes or positions could potentially block flow more severely, leading to substantial pressure drops in the left coronary arteries.

Subjects

Author keywords

computational fluid dynamicscoronarystent thrombosis

Authors and institutions

  • Department of Mechanical Engineering, Pontifical Catholic University of Rio de Janeiro, Rio de Janeiro 22451-900, RJ, Brazil.
  • Fit Center-Clínica de Performance Humana, Niterói 24220-331, RJ, Brazil.
  • Hospital do Coração-HCor, São Paulo 04004-030, SP, Brazil.

ORCID profiles: Ivan Fernney Ibanez Aguilar, Pedro Soares Teixeira, Adriano Camargo Carneiro, Bruno Alvares de Azevedo, Angela Ourivio Nieckele, Carlos Eduardo Rochitte

Cite

APA

Ibanez Aguilar, I. F., Beltrão Paraizo, M., Soares Teixeira, P., Camargo Carneiro, A., Alvares de Azevedo, B., Nieckele, A. O., & Rochitte, C. E. (2026). A Study of Coronary Stent Thrombosis Growth. Bioengineering, 13(7), 779. https://doi.org/10.3390/bioengineering13070779

Vancouver

Ibanez Aguilar IF, Beltrão Paraizo M, Soares Teixeira P, Camargo Carneiro A, Alvares de Azevedo B, Nieckele AO, et al. A Study of Coronary Stent Thrombosis Growth. Bioengineering. 2026;13(7):779. doi:10.3390/bioengineering13070779

BibTeX
@article{IbanezAguilar2026,
  author = {Ibanez Aguilar, Ivan Fernney and Beltrão Paraizo, Marcos and Soares Teixeira, Pedro and Camargo Carneiro, Adriano and Alvares de Azevedo, Bruno and Nieckele, Angela Ourivio and Rochitte, Carlos Eduardo},
  title = {A Study of Coronary Stent Thrombosis Growth},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {7},
  pages = {779},
  doi = {10.3390/bioengineering13070779}
}