A Study of Coronary Stent Thrombosis Growth
Bioengineering · 07.07.2026 · 13 (7) · 779
DOI: 10.3390/bioengineering13070779 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 07.07.2026. Verfasst von Ivan Fernney Ibanez Aguilar und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, Pontifical Catholic University of Rio de Janeiro, Rio de Janeiro 22451-900, RJ, Brazil. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
In this study, the impact of thrombus growth at the left coronary ostium—resulting from a misplaced stent—was investigated through numerical simulations. Blood flow from the aortic root into the coronary arteries was modeled for several potential thrombus sizes to assess changes in pressure and shear stress caused by thrombus development. The results showed that even in the most severe case examined (with the thrombus growing three times its original size), the obstruction to the coronary cross-sectional area was not sufficient to produce a significant pressure drop or wall shear stress increases. This outcome is primarily attributed to the thrombus’s location at the tip of the stent; even with considerable elongation, it did not significantly narrow the coronary artery’s lumen. While the specific configurations studied did not result in critical obstruction, it is important to note that other thrombus shapes or positions could potentially block flow more severely, leading to substantial pressure drops in the left coronary arteries.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Mechanical Engineering, Pontifical Catholic University of Rio de Janeiro, Rio de Janeiro 22451-900, RJ, Brazil.
- Fit Center-Clínica de Performance Humana, Niterói 24220-331, RJ, Brazil.
- Hospital do Coração-HCor, São Paulo 04004-030, SP, Brazil.
ORCID-Profile: Ivan Fernney Ibanez Aguilar, Pedro Soares Teixeira, Adriano Camargo Carneiro, Bruno Alvares de Azevedo, Angela Ourivio Nieckele, Carlos Eduardo Rochitte
Zitieren
APA
Ibanez Aguilar, I. F., Beltrão Paraizo, M., Soares Teixeira, P., Camargo Carneiro, A., Alvares de Azevedo, B., Nieckele, A. O., & Rochitte, C. E. (2026). A Study of Coronary Stent Thrombosis Growth. Bioengineering, 13(7), 779. https://doi.org/10.3390/bioengineering13070779
Vancouver
Ibanez Aguilar IF, Beltrão Paraizo M, Soares Teixeira P, Camargo Carneiro A, Alvares de Azevedo B, Nieckele AO, et al. A Study of Coronary Stent Thrombosis Growth. Bioengineering. 2026;13(7):779. doi:10.3390/bioengineering13070779
BibTeX
@article{IbanezAguilar2026,
author = {Ibanez Aguilar, Ivan Fernney and Beltrão Paraizo, Marcos and Soares Teixeira, Pedro and Camargo Carneiro, Adriano and Alvares de Azevedo, Bruno and Nieckele, Angela Ourivio and Rochitte, Carlos Eduardo},
title = {A Study of Coronary Stent Thrombosis Growth},
journal = {Bioengineering},
year = {2026},
volume = {13},
number = {7},
pages = {779},
doi = {10.3390/bioengineering13070779}
}
EUVIMED