EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Tissue Structure as a Primary Factor Influencing Vascular Sealing: Results of an Ex Vivo Study on Porcine Carotid Arteries

Andreas Kirschbaum, Dimitri Raico, Florian Kirschbaum, Moritz Jesinghaus, Nikolas Mirow iD

Bioengineering · 24.06.2026 · 13 (7) · 719

DOI: 10.3390/bioengineering13070719 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 24.06.2026. Verfasst von Andreas Kirschbaum und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Visceral, Thoracic, and Vascular Surgery, University Hospital Giessen and Marburg (UKGM), Marburg Campus, Baldingerstrasse 1, D-35043 M…. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Bipolar vessel sealing systems are widely used in surgery, yet their effectiveness varies depending on the histological composition of the target vessel. In particular, the influence of elastin on seal stability is not well understood. Porcine carotid arteries, which show a pronounced proximal–distal elastin gradient, provide an ideal model for systematic analysis. In this study, fresh porcine carotid arteries were divided into three segments based on vessel diameter (7 mm). Histological EvG staining was used to quantify elastin and collagen content. All vessels (n = 8 per group) were sealed using a bipolar marSeal® 5 plus device, followed by burst pressure testing and peel force measurements. Elastin content increased significantly from peripheral to central segments (9% → 25% → 42%; p < 0.001), while collagen content remained constant (22 ± 2%). In parallel, seal stability decreased markedly: burst pressures dropped from 723 mmHg to 240 mmHg and to 31.5 mmHg (p < 0.001). Peel forces showed the same trend (1.75 ± 0.07 N → 0.65 ± 0.03 N → 0.26 ± 0.11 N; p < 0.001). Wall thickness showed no proportional relationship to seal quality. Interestingly, the sealing performance of bipolar systems seems to be greatly influenced by the histological structure of the vessel wall. A high elastin content—rising from 9% to 42% along the carotid artery—was associated with a reduction in burst pressure and peel strength. These findings highlight the need to consider tissue composition when selecting sealing methods and support the development of adaptive energy delivery technologies.

Schlagwörter

Stichwörter der Autor:innen

bipolar vessel sealingburst pressureelastin contentex vivo studyhistological vessel structureporcine carotid artery

Autor:innen und Institutionen

  • Department of Visceral, Thoracic, and Vascular Surgery, University Hospital Giessen and Marburg (UKGM), Marburg Campus, Baldingerstrasse 1, D-35043 Marburg, Germany.
  • General Surgery, Eichhof Hospital, Eichhofstrasse 1, D-36341 Lauterbach, Germany.
  • Department of Human Medicine, Albert Ludwig University of Freiburg, Breisacher Strasse 153, D-79110 Freiburg im Breisgau, Germany.
  • Institute of Pathology, University Hospital Giessen and Marburg (UKGM), Marburg Campus, Baldingerstrasse 1, D-35043 Marburg, Germany.
  • Department of Human Medicine, Philipps University of Marburg, Baldingerstrasse 1, D-35043 Marburg, Germany.

ORCID-Profile: Nikolas Mirow

Zitieren

APA

Kirschbaum, A., Raico, D., Kirschbaum, F., Jesinghaus, M., & Mirow, N. (2026). Tissue Structure as a Primary Factor Influencing Vascular Sealing: Results of an Ex Vivo Study on Porcine Carotid Arteries. Bioengineering, 13(7), 719. https://doi.org/10.3390/bioengineering13070719

Vancouver

Kirschbaum A, Raico D, Kirschbaum F, Jesinghaus M, Mirow N. Tissue Structure as a Primary Factor Influencing Vascular Sealing: Results of an Ex Vivo Study on Porcine Carotid Arteries. Bioengineering. 2026;13(7):719. doi:10.3390/bioengineering13070719

BibTeX
@article{Kirschbaum2026,
  author = {Kirschbaum, Andreas and Raico, Dimitri and Kirschbaum, Florian and Jesinghaus, Moritz and Mirow, Nikolas},
  title = {Tissue Structure as a Primary Factor Influencing Vascular Sealing: Results of an Ex Vivo Study on Porcine Carotid Arteries},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {7},
  pages = {719},
  doi = {10.3390/bioengineering13070719}
}
Kostenlos starten →