EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A Multi-Channel AM-TMAS Driving System Based on Amplitude-Modulated Sine Waves

Yiheng Shi, Ze Li, Ruixu Liu, Xiyang Zhang iD, Mingpeng Wang, Ren Ma iD, Tao Yin, Xiaoqing Zhou, Zhipeng Liu

Bioengineering · 31.03.2026 · 13 (4) · 405

DOI: 10.3390/bioengineering13040405 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 31.03.2026. Verfasst von Yiheng Shi und 8 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Advanced Medical Materials and Devices, Tianjin 300192, China. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, CAMS Innovation Fund for Medical Sciences, Natural Science Foundation of Tianjin.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Selectively modulating specific brain-rhythm bands with physical stimuli helps both to reveal neural mechanisms and to provide non-pharmacological treatment avenues for brain disorders. This study proposes and implements a multi-channel transcranial magneto-acoustic stimulation driving system based on amplitude-modulated (AM) sine waves (AM-TMAS) intended to supply a reliable hardware platform for noninvasive, focal low-frequency rhythmic electrical stimulation of deep-brain structures. The driving system implements a 64-channel AM module based on an FPGA plus high-speed DACs. Multi-channel precision is achieved via a unified high-speed clock and a global UPDATE trigger. To overcome the large separation between envelope and carrier frequencies, we developed a high-fidelity AM waveform generation method based on DDS + LUT + envelope multiplication. The algorithm first centers the carrier samples to preserve waveform symmetry, then applies LUT-based envelope coefficients and fixed-point envelope multiplication, enabling high-precision AM outputs with carrier frequencies from 100 kHz to 2 MHz and envelope frequencies from 0.1 Hz to 100 kHz. We tested the system’s rhythmic multi-channel AM output performance across frequencies and also measured magneto-acoustic-coupled rhythmic electrical signals produced by the AM-TMAS driving setup. Any single channel reliably produced high-fidelity AM waveforms with a 500 kHz carrier and 8 Hz/40 Hz envelopes; the measured carrier was 499.998 kHz with excellent frequency stability. Both envelope and carrier frequencies are flexibly tunable. At the nominal 500 kHz carrier, envelope fidelity was further quantified: the extracted envelopes achieved NRMSEs of 1.0795% (8 Hz) and 1.9212% (40 Hz), confirming high-fidelity AM synthesis. Under a 0.3 T static magnetic field, the AM-TMAS driving system generated rhythmic electrical responses in physiological saline that carried the expected 40 Hz envelope. The proposed AM-TMAS driver achieves high accuracy in AM waveform generation and robust multi-channel performance, and—when combined with an external static magnetic field—can produce rhythmically modulated magneto-acoustic electrical stimulation. This platform provides a practical technical tool for brain-function research and the development of rhythm-targeted neuromodulation therapies.

Schlagwörter

Stichwörter der Autor:innen

amplitude modulationdeep brain stimulationfield-programmable gate array (FPGA)multi-channel synchronizationrhythmic neural stimulationtranscranial magneto-acoustic stimulation (TMAS)

Autor:innen und Institutionen

  • State Key Laboratory of Advanced Medical Materials and Devices, Tianjin 300192, China.
  • Tianjin Key Laboratory of Neuromodulation and Neurorepair, Tianjin 300192, China.
  • Institute of Biomedical Engineering, Peking Union Medical College, Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Institute of Biomedical Engineering, Tianjin Institutes of Health Science, Tianjin 300192, China.

ORCID-Profile: Xiyang Zhang, Ren Ma

Förderung

  • National Natural Science Foundation of China
  • CAMS Innovation Fund for Medical Sciences
  • Natural Science Foundation of Tianjin
  • Fundamental Research Funds for the Central Universities of Peking Union Medical College
  • State Key Laboratory of Advanced Medical Materials and Devices Research Grant
  • National Natural Science Foundation of China https://ror.org/01h0zpd94

Zitieren

APA

Shi, Y., Li, Z., Liu, R., Zhang, X., Wang, M., Ma, R., Yin, T., Zhou, X., & Liu, Z. (2026). A Multi-Channel AM-TMAS Driving System Based on Amplitude-Modulated Sine Waves. Bioengineering, 13(4), 405. https://doi.org/10.3390/bioengineering13040405

Vancouver

Shi Y, Li Z, Liu R, Zhang X, Wang M, Ma R, et al. A Multi-Channel AM-TMAS Driving System Based on Amplitude-Modulated Sine Waves. Bioengineering. 2026;13(4):405. doi:10.3390/bioengineering13040405

BibTeX
@article{Shi2026,
  author = {Shi, Yiheng and Li, Ze and Liu, Ruixu and Zhang, Xiyang and Wang, Mingpeng and Ma, Ren and Yin, Tao and Zhou, Xiaoqing and Liu, Zhipeng},
  title = {A Multi-Channel AM-TMAS Driving System Based on Amplitude-Modulated Sine Waves},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {4},
  pages = {405},
  doi = {10.3390/bioengineering13040405}
}
Kostenlos starten →