Fachartikel in der Zeitschrift Bioengineering, online erschienen am 24.03.2026. Verfasst von Seongho Woo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physical Medicine and Rehabilitation, Daegu Fatima Hospital, Daegu-si 41121, Republic of Korea. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea (NRF), Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), the ‘Junior Faculty Research Support Grant’ at Changwon National University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Cervical spondylosis is a common cause of spinal cord dysfunction, and anterior cervical discectomy and fusion (ACDF) is widely employed when conservative treatment fails. Conventional implant systems such as the cervical cage with plate (CCP) and zero-profile stand-alone cage (ZPSC) are commonly used to enhance spinal stability and promote fusion, but they are associated with complications including dysphagia and adjacent segment degeneration. To address these limitations, a novel flexible plate cage system (FPCS) has been developed to optimize biomechanical performance while minimizing surgical risk. In this study, a finite element model of the C3–T1 cervical spine was constructed to simulate ACDF at the C5–C6 level using CCP, ZPSC, and FPCS implants. Under standardized loading conditions, von Mises stress was analyzed in the bone, intervertebral disc, endplates, cage, and screws, using the mean of the top 5% stress values to ensure accuracy. All surgical models showed increased stress compared to the intact reference spine. The ZPSC model exhibited the highest stress in the cage and screws, suggesting a more concentrated load path. The CCP model showed a more evenly distributed stress profile, particularly affecting the inferior adjacent segment. The FPCS model demonstrated moderate cage stress, reduced screw stress, and the highest plate stress, indicating a design that effectively redirects mechanical load from the screw-bone interface toward the anterior plate. This may be related to the unique structural configuration of the FPCS, which secures screws horizontally into the anterior vertebral body without penetrating the endplates. These findings suggest that the FPCS may offer a biomechanically favorable alternative to existing ACDF implants.

Schlagwörter

Stichwörter der Autor:innen

anterior cervical discectomy and fusion (ACDF)cervical cage with plate (CCP)flexible plate cage system (FPCS)zero-profile stand-alone cage (ZPSC)

Autor:innen und Institutionen

  • Department of Physical Medicine and Rehabilitation, Daegu Fatima Hospital, Daegu-si 41121, Republic of Korea.
  • Department of Physical Medicine and Rehabilitation, Haengbokhan Rehabilitation Medicine Clinic, Daegu-si 42174, Republic of Korea.
  • Department of Smart Ocean Mobility Engineering, Changwon National University, Changwon-si 51140, Republic of Korea.

ORCID-Profile: Jong-Moon Hwang, Sungwook Kang

Förderung

  • National Research Foundation of Korea (NRF)
  • Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI)
  • the ‘Junior Faculty Research Support Grant’ at Changwon National University
  • Chung Hie Oh & Jin-Sang Chung research grant of Korean Academy of Rehabilitation Medicine

Zitieren

APA

Woo, S., Koo, W. M., Park, K., Hwang, J. M., & Kang, S. (2026). Evaluation of a Novel Flexible Cage System for C5–C6 Fixation: A Finite Element Study Against Conventional ACDF Implants. Bioengineering, 13(4), 375. https://doi.org/10.3390/bioengineering13040375

Vancouver

Woo S, Koo WM, Park K, Hwang JM, Kang S. Evaluation of a Novel Flexible Cage System for C5–C6 Fixation: A Finite Element Study Against Conventional ACDF Implants. Bioengineering. 2026;13(4):375. doi:10.3390/bioengineering13040375

BibTeX
@article{Woo2026,
  author = {Woo, Seongho and Koo, Won Mo and Park, Kinam and Hwang, Jong-Moon and Kang, Sungwook},
  title = {Evaluation of a Novel Flexible Cage System for C5–C6 Fixation: A Finite Element Study Against Conventional ACDF Implants},
  journal = {Bioengineering},
  year = {2026},
  volume = {13},
  number = {4},
  pages = {375},
  doi = {10.3390/bioengineering13040375}
}
Kostenlos starten →