Design and Evaluation of Stand-to-Sit and Sit-to-Stand Control Protocols for a HIP–Knee–Ankle–Foot Prosthesis with a Motorized Hip Joint
Bioengineering · 2025-12-31 · 13 (1) · 48
DOI: 10.3390/bioengineering13010048 ↗At a glance
Journal article in Bioengineering, published online on 2025-12-31. Written by Farshad Golshan and 4 co-authors. Institution: Department of Mechanical Engineering, University of Ottawa, Ottawa, ON K1N 6N5, Canada. Cited 2 times so far (Crossref); 36 references. The full text is freely available (open access, CC BY 4.0). Funded by: Mitacs, Natural Sciences and Engineering Research Council of Canada, Mitacs and Össur.
Abstract (original language)
Background
Sitting and standing with conventional hip–knee–ankle–foot (HKAF) prostheses are demanding tasks for hip disarticulation (HD) amputees due to the passive nature of current prosthetic hip joints that cannot assist with moment generation. This study developed a sitting and standing control strategy for a motorized hip joint and evaluated whether providing active assistance reduces the intact side demand of these activities.
Methods
A dedicated control strategy was developed and implemented for a motorized hip prosthesis (Power Hip) compatible with existing prosthetic knees, feet, and sockets. One HD participant was trained to perform sitting and standing tasks using the Power Hip. Its performance was compared with the participant’s prescribed passive HKAF prosthesis through measurements of ground reaction forces (GRFs), joint moments, and activity durations. GRFs were collected using force plates, kinematics were captured via Theia3D markerless motion capture, and joint moments were computed in Visual3D.
Results
The Power Hip enabled more symmetric limb loading and faster stand-to-sit transitions (1.22 ± 0.08 s vs. 2.62 ± 0.41 s), while slightly prolonging sit-to-stand (1.69 ± 0.49 s vs. 1.22 ± 0.40 s) compared to the passive HKAF. The participant exhibited reduced intact-side loading impulses during stand-to-sit (4.97 ± 0.78 N∙s/kg vs. 15.06 ± 2.90 N∙s/kg) and decreased reliance on upper-limb support. Hip moment asymmetries between the intact and prosthetic sides were also reduced during both sit-to-stand (−0.18 ± 0.09 N/kg vs. −0.69 ± 0.67 N/kg) and stand-to-sit transitions (0.77 ± 0.20 N/kg vs. 2.03 ± 0.58 N/kg).
Conclusions
The prototype and control strategy demonstrated promising improvements in sitting and standing performance compared to conventional passive prostheses, reducing the physical demand on the intact limb and upper body.
Subjects
Author keywords
Authors and institutions
- Department of Mechanical Engineering, University of Ottawa, Ottawa, ON K1N 6N5, Canada.
- Faculty of Medicine, University of Ottawa, Ottawa, ON K1H 8M5, Canada.
- Ottawa Hospital Research Institute, Ottawa, ON K1H 8L6, Canada.
- Nielen Orthotics and Prosthetics, Ottawa, ON K2A 2C5, Canada.
ORCID profiles: Farshad Golshan, Natalie Baddour, Hossein Gholizadeh, Edward D Lemaire
Funding
- Mitacs
- Natural Sciences and Engineering Research Council of Canada
- Mitacs and Össur
Cite
APA
Golshan, F., Baddour, N., Gholizadeh, H., Nielen, D., & Lemaire, E. D. (2025). Design and Evaluation of Stand-to-Sit and Sit-to-Stand Control Protocols for a HIP–Knee–Ankle–Foot Prosthesis with a Motorized Hip Joint. Bioengineering, 13(1), 48. https://doi.org/10.3390/bioengineering13010048
Vancouver
Golshan F, Baddour N, Gholizadeh H, Nielen D, Lemaire ED. Design and Evaluation of Stand-to-Sit and Sit-to-Stand Control Protocols for a HIP–Knee–Ankle–Foot Prosthesis with a Motorized Hip Joint. Bioengineering. 2025;13(1):48. doi:10.3390/bioengineering13010048
BibTeX
@article{Golshan2025,
author = {Golshan, Farshad and Baddour, Natalie and Gholizadeh, Hossein and Nielen, David and Lemaire, Edward D},
title = {Design and Evaluation of Stand-to-Sit and Sit-to-Stand Control Protocols for a HIP–Knee–Ankle–Foot Prosthesis with a Motorized Hip Joint},
journal = {Bioengineering},
year = {2025},
volume = {13},
number = {1},
pages = {48},
doi = {10.3390/bioengineering13010048}
}
EUVIMED