EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Dynamic Cerebral Perfusion Electrical Impedance Tomography: A Neuroimaging Technique for Bedside Cerebral Perfusion Monitoring During Mannitol Dehydration

Weice Wang iD, Lihua Hou, Canhua Xu iD, Mingxu Zhu iD, Yitong Guo iD, Rong Zhao, Weixun Duan, Yu Wang iD, Zhenxiao Jin iD, Xuetao Shi iD

Bioengineering · 31.10.2025 · 12 (11) · 1187

DOI: 10.3390/bioengineering12111187 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 31.10.2025. Verfasst von Weice Wang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Shaanxi Provincial Key Laboratory of Bioelectromagnetic Detection and Intelligent Perception, Air Force Medical…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China-General Program, Cross-integration Project of Xijing Hospital, Key Research and Development Projects of the Science and Technology Committee.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Mannitol dehydration is routinely used to prevent and treat cerebral damage after total aortic arch replacement (TAAR), but existing neuroimaging technologies cannot achieve bedside real-time quantitative assessment of its impact on cerebral perfusion in different patients. This study applied dynamic cerebral perfusion electrical impedance tomography (DCP-EIT), a non-invasive neuroimaging technique, for bedside cerebral perfusion monitoring in TAAR patients during dehydration. Seventeen patients with normal neurological function and nineteen with neurological dysfunction (ND) were enrolled. The variation patterns and differences in perfusion impedance, images, and the relative ratios (RY) of mean perfusion velocity (MV), height of systolic wave (Hs), inflow volume velocity (IV), and angle between the ascending branch and baseline (Aab) were analyzed. Results showed DCP-EIT could visualize cerebral perfusion changes, with detected poorly perfused regions showing good consistency with ischemic areas identified by computed tomography (CT). RY of normal patients fluctuated around 0.97–1.04, with no significant difference from baseline. RY of ND patients peaked at 14–20 min after dehydration and remained higher than baseline even at 100 min (p < 0.001). DCP-EIT holds potential to optimize individualized cerebral protection strategies for other cerebral damage scenarios and neurocritical care.

Schlagwörter

Stichwörter der Autor:innen

bedside monitoringbraincerebral perfusion imagingelectrical impedance tomographyneuroimagingtotal aortic arch replacement

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Shaanxi Provincial Key Laboratory of Bioelectromagnetic Detection and Intelligent Perception, Air Force Medical University, Xi'an 710032, China.
  • Department of Cardiovascular Surgery, Xijing Hospital, Air Force Medical University, Xi'an 710032, China.
  • Institute of Medical Research, Northwestern Polytechnical University, Xi'an 710072, China.
  • Department of Ultrasound Diagnosis, Tangdu Hospital, Air Force Medical University, Xi'an 710038, China.

ORCID-Profile: Weice Wang, Canhua Xu, Mingxu Zhu, Yitong Guo, Yu Wang, Zhenxiao Jin, Xuetao Shi

Förderung

  • National Natural Science Foundation of China-General Program
  • Cross-integration Project of Xijing Hospital
  • Key Research and Development Projects of the Science and Technology Committee
  • National Key Research and Development Program
  • Health Development Promotion Project
  • Interdisciplinary Integration Special Project of Air Force Medical University

Zitieren

APA

Wang, W., Hou, L., Xu, C., Zhu, M., Guo, Y., Zhao, R., Duan, W., Wang, Y., Jin, Z., & Shi, X. (2025). Dynamic Cerebral Perfusion Electrical Impedance Tomography: A Neuroimaging Technique for Bedside Cerebral Perfusion Monitoring During Mannitol Dehydration. Bioengineering, 12(11), 1187. https://doi.org/10.3390/bioengineering12111187

Vancouver

Wang W, Hou L, Xu C, Zhu M, Guo Y, Zhao R, et al. Dynamic Cerebral Perfusion Electrical Impedance Tomography: A Neuroimaging Technique for Bedside Cerebral Perfusion Monitoring During Mannitol Dehydration. Bioengineering. 2025;12(11):1187. doi:10.3390/bioengineering12111187

BibTeX
@article{Wang2025,
  author = {Wang, Weice and Hou, Lihua and Xu, Canhua and Zhu, Mingxu and Guo, Yitong and Zhao, Rong and Duan, Weixun and Wang, Yu and Jin, Zhenxiao and Shi, Xuetao},
  title = {Dynamic Cerebral Perfusion Electrical Impedance Tomography: A Neuroimaging Technique for Bedside Cerebral Perfusion Monitoring During Mannitol Dehydration},
  journal = {Bioengineering},
  year = {2025},
  volume = {12},
  number = {11},
  pages = {1187},
  doi = {10.3390/bioengineering12111187}
}
Kostenlos starten →