Mandibular Titanium Miniplates Change the Biomechanical Behaviour of the Mandible in the Case of Facial Trauma: A Three-Dimensional Finite Element Analysis
Bioengineering · 22.08.2023 · 10 (9) · 994
DOI: 10.3390/bioengineering10090994 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bioengineering, online erschienen am 22.08.2023. Verfasst von Nicolas Graillon und 5 weiteren Autor:innen. Beteiligte Einrichtung: Laboratoire de Bioméchanique Appliquée (LBA), Gustave Eiffel University/Aix-Marseille University, 13015 Marseille, France. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Our study aimed to compare the biomechanical behaviour of mandibles with or without titanium miniplates when subjected to an impact after bone healing using a finite element model (FEM) of the human mandible. We simulated mandibular trauma on an FEM of a human mandible carrying or not two parasymphyseal miniplates and applying a concentrated force of 2000 N to four different areas, including the insertion area, the area straddling the edge of the miniplates and the adjacent bone, at a distance from the miniplates on the symphysis, and on the basilar border of the mandible below the miniplates. Then, we compared the Von Mises stress distributions between the two models. In the case of an impact on the miniplates, the maximum Von Mises stress occurred in two specific areas, on the cortical bone at the posterior border of the two miniplates at a distance from the impact, while in the model without miniplates, the Von Mises stresses were homogenously distributed in the impact area. The presence of titanium miniplates in the case of trauma affects the biomechanical behaviour of the mandible and could cause more complex fractures. We recommend informing patients of this potential risk.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratoire de Bioméchanique Appliquée (LBA), Gustave Eiffel University/Aix-Marseille University, 13015 Marseille, France.
- Department of Oral and Maxillofacial Surgery/Assistance Publique-Hopitaux de Marseille (APHM), Conception University Hospital, 13005 Marseille, France.
- Maxillo-Facial Surgery, Facial Plastic Surgery, Stomatology and Oral Surgery, Hospices Civils de Lyon, Lyon-Sud Hospital-Claude-Bernard Lyon 1 University, 69310 Pierre-Benite, France.
ORCID-Profile: Nicolas Graillon, Yves Godio-Raboutet, Andrea Varazzani, Lionel Thollon
Zitieren
APA
Graillon, N., Foletti, J. M., Godio-Raboutet, Y., Guyot, L., Varazzani, A., & Thollon, L. (2023). Mandibular Titanium Miniplates Change the Biomechanical Behaviour of the Mandible in the Case of Facial Trauma: A Three-Dimensional Finite Element Analysis. Bioengineering, 10(9), 994. https://doi.org/10.3390/bioengineering10090994
Vancouver
Graillon N, Foletti JM, Godio-Raboutet Y, Guyot L, Varazzani A, Thollon L. Mandibular Titanium Miniplates Change the Biomechanical Behaviour of the Mandible in the Case of Facial Trauma: A Three-Dimensional Finite Element Analysis. Bioengineering. 2023;10(9):994. doi:10.3390/bioengineering10090994
BibTeX
@article{Graillon2023,
author = {Graillon, Nicolas and Foletti, Jean-Marc and Godio-Raboutet, Yves and Guyot, Laurent and Varazzani, Andrea and Thollon, Lionel},
title = {Mandibular Titanium Miniplates Change the Biomechanical Behaviour of the Mandible in the Case of Facial Trauma: A Three-Dimensional Finite Element Analysis},
journal = {Bioengineering},
year = {2023},
volume = {10},
number = {9},
pages = {994},
doi = {10.3390/bioengineering10090994}
}
EUVIMED